DCS Industrial control system
NameDescriptionContent
NEW CENTER
Current Location:

ABBEasyLine EL3000 Series

From:ABB | Author:LIAO | Time :2025-07-29 | 643 Browse: | 🔊 Click to read aloud ❚❚ | Share:

Measurement made easy So smart, they’re simple

— 

A wide variety of measurement technology Ů Detectors for multiple process and emission monitoring applications Ů Up to five sample components in one gas analyzer Ů Suitable for measuring flammable gases Ů Version in Category 3G for measuring non-flammable gases Ů Performance-tested for emission monitoring according to EN 15267 Ů QAL3 monitoring according to EN 14181 (optional)

— 

Straightforward handling Ů Automatic calibration including control of pump and solenoid valves Ů Simplified calibration with air or integral calibration cells eliminating the need for test gas cylinders Ů Self-monitoring function indicates when maintenance is required

— 

Application-oriented design Ů Housing design for 19-inch rack mounting or wall mounting Ů Integrated pneumatics module in model EL3020 (optional) Ů Ethernet, Modbus and PROFIBUS interfaces Ů Configurable analog outputs and digital inputs/outputs

 User-friendly operation Ů Simple operator interface Ů Clear-text status messages Ů Configuration of seldom required functions with configuration software


Overview of the gas analyzers Measuring technology (analyzers)

The following analyzers are available for selection:

Uras26 infrared photometer for the measurement of infrared-active gas components, e.g. CO, NO, SO2 

• Limas23 ultraviolet photometer for the measurement of NO, NO2 and SO2 • Magnos206 oxygen analyzer for the measurement of O2 in process gas or in N2 

• Magnos28 oxygen analyzer for the measurement of O2 in process gas or in N2 

• Magnos27 oxygen analyzer for the measurement of O2 in flue gas or in N2 

• ZO23 trace oxygen analyzer for the measurement of O2 in pure gases (N2, CO2, Ar)

• Caldos27 thermal conductivity analyzer for the measurement of binary gas mixtures with different thermal conductivity, e.g. Ar in O2, H2 in Ar, CH4 in N2 

• Fidas24 flame-ionization detector for the measurement of hydrocarbons 

• Electrochemical oxygen sensor for the measurement of O2 Magnos206 and Magnos28 can also be used in combination with Uras26 or Limas23. Magnos27 and Caldos27 can also be used in combination with Uras26. Fidas24 and ZO23 cannot be used in combination with one of the other analyzers. The electrochemical oxygen sensor can only be used in combination with Uras26 or Limas23. Two electrochemical oxygen sensors can be used in combination with Uras26 with separate gas paths (only in model EL3020). Each analyzer has one physical measurement range per sample component. A section of the physical measurement range can be mapped to the current output (analog output) by on-site configuration. Calibration is always executed in the physical measurement range. The permissible measurement range limits are given by the specification of the smallest and largest measurement ranges for the individual analyzers. A total of up to five measurement components can be measured with one gas analyzer.

Calibration

Calibration can be performed automatically or manually or externally controlled via the Modbus. Automatic calibration – for all sample components together – is normally started on a cyclically time-controlled basis; it can also be started by an external control signal or via the Modbus as well as manually on the display and operator control unit of the gas analyzer. Normally, simplified calibration methods with the built-in calibration cells or the so-called single-point calibration are used for automatic calibration. If calibration with test gases is required the control of solenoid valves for switching on five test gases, z

Concept of operation

The functions required in normal operation are operated and configured directly on the gas analyzer. The functions which are only seldom required, e.g. during startup, are configured offline using the configuration software ECT (“EasyLine Configuration Tool” on the enclosed DVD-ROM) and then loaded into the gas analyzer.

QAL3 monitoring

QAL3 monitoring is available as an option in the gas analyzer. It is used to fulfill the requirements according to EN 14181 for storage and analysis of device adjustment data. The QAL3 monitoring option features the following functions: 

• Automatic acquisition, verification and documentation of drift and precision at zero and reference points 

• Reporting via CUSUM and Shewhart control charts 

• QAL3 data storage in the gas analyzer (maximum 1 year) 

• QAL3 data display and read-out as well as parameter setting via web browser 

• Status messages on deviations beyond requirements 

• Data export for further processing with spreadsheet programs

Electrical interfaces

The electrical interfaces for the output of measured values and communication with external systems include 

• The integrated Ethernet-10/100BASE-T interface for – gas analyzer configuration with configuration software ECT – QAL3 data transfer if the QAL3 monitoring option is integrated in the gas analyzer – data transmission with Modbus TCP/IP protocol (measured values, status signals and control signals) as well as the integrated I/O modules depending on the functional range and order  Profibus module with one RS485 and one MBP interface (also according to VDI 4201 Part 2),  Modbus module with one RS485 and one RS232 interface (also according to VDI 4201 Part 3),  Digital I/O module with four digital inputs and four digital outputs,  2-way analog output module with two analog outputs,  4-way analog output module with four analog outputs.

Integral gas feed

The integral gas feed (optional in model EL3020) is available in two versions. It includes 

• either the solenoid valve, pump, coarse filter, capillary tube and flow sensor modules 

• or a flow sensor module.

Housing design

The housing for the EL3020 gas analyzer model is designed as 19-inch housing with 3 height units (4 height units with Magnos27) and degree of protection IP20 (IP40 with version for emission monitoring). The housing for the EL3040 gas analyzer model is designed as wall-mount housing with degree of protection IP65.


Dynamic response

Warm-up time Approx. 30 minutes without thermostat; approx. 2 hours with thermostat

90% response time T90 ≤ 2.5 sec for sample gas flow = 60 l/h and electronic T90 time (static/dynamic) = 5/0 sec

Calibration

Zero-point calibration With inert gas, e.g. nitrogen, or with ambient air that is free of the sample component. End-point calibration With gas-filled calibration cells (optional) or with test gas mixtures. It is recommended to verify the calibration cell set values once a year.

Materials in contact with the sample medium Analyzer (sample cells) Tubing: aluminum; windows

Tubing: aluminum; windows: CaF2 or BaF2; connectors: stainless steel 1.4305 (SAE 303) Gas lines and connectors Gas lines: FPM hoses or PTFE tubes; connectors: stainless steel 1.4305 (SAE 303); solenoid valve (option in model EL3020): PVDF. When flammable components are present in the sample gas: Gas lines: stainless steel tubes 1.4571 (SAE 316Ti); connectors: stainless steel 1.4305 (SAE 303).

Gas connections

See page 27 and page 28

Sample gas inlet conditions

The analyzer must not be used for measurement of ignitable gas/air or gas/oxygen mixtures.

Temperature

The sample gas dew point should be at least 5 °C below the temperature throughout the sample gas path. Otherwise a sample gas cooler or condensate trap is required.

Pressure

The analyzer is operated under atmospheric pressure; the sample gas outlet is open to atmosphere. Internal pressure drop < 5 hPa with standard flow rate 60 l/h. Permissible absolute pressure range: 800–1250 hPa. Operation under lower absolute pressure (e.g. at altitudes above 2000 m) on request. Overpressure in the sample cell max. 500 hPa.

Flow rate

20–100 l/h

Corrosive gases

Highly corrosive associated gas components, e.g. chlorine (Cl2) and hydrogen chloride (HCl), as well as gases or aerosols containing chlorine must be cooled or undergo prior absorption.

Flammable gases

In the version with gas lines and connectors made of stainless steel the analyzer is suitable for measuring flammable gases in general purpose environment (see page 22).


Dynamic response

Warm-up time

< 1 hour

90% response time

T90 approx. 4 sec at a sample gas flow of 90 l/h and electronic

T90 time (static/dynamic) = 3/0 sec, gas change from nitrogen

to air

Calibration

Zero-point calibration

With oxygen-free process gas or substitute gas

End-point calibration

With process gas with a known oxygen concentration or a substitute gas such as dried air

Single-point calibration

Zero-point calibration with any oxygen concentration, e.g. with

nitrogen or ambient air, processed through a cooler or H2O absorber.

Pressure correction by means of pressure sensor is recommended for single-point calibration with air.

Depending on the measurement task involved, the zero- and endpoints should be verified periodically (recommendation: once a

year).

Calibration of measurement ranges with suppressed zero-point

Highly suppressed measurement ranges (≥95–100 vol.% O2)

should only be calibrated with test gases with concentrations in

the selected measurement range.

Materials in contact with the sample medium

Analyzer

Sample chamber (direct connection): stainless steel 1.4305

(SAE 303), glass, platinum, rhodium, epoxy resin; seals: FPM (Fluorocarbon rubber), PEEK, FFKM

Gas connectors of the solenoid valve (option in model EL3020)

PVDF

Gas connections

See page 30

Sample gas inlet conditions

The analyzer must not be used for measurement of ignitable

gas/air or gas/oxygen mixtures.

Temperature

The sample gas dew point should be at least 5 °C below the temperature throughout the sample gas path. Otherwise a sample

gas cooler or condensate trap is required. Water vapor content

variations cause volume errors.

Pressure

Operation under atmospheric pressure: The sample gas outlet is

open to atmosphere. Internal pressure drop < 5 hPa with standard flow rate 60 l/h. Permissible absolute pressure range: 800–

1250 hPa. Operation under lower absolute pressure (e.g. at altitudes above 2000 m) on request.

Operation under elevated pressure: A pressure sensor is required

for pressure influence compensation.

Absolute pressure ≤ 1250 hPa: An optional internal pressure sensor can be connected to the sample gas path.

Absolute pressure ≥ 1250 hPa: An external pressure sensor must

be connected to the sample gas path.

The analyzer is function-tested for 5000 hPa internal pressure

without damage.

Flow rate

30–90 l/h

Corrosive gases

Consultation with ABB Analytical is required if the sample gas

contains Cl2, HCl, HF or other corrosive components.

The AO2000-Magnos206 analyzer should be used if the sample

gas contains NH3.

Flammable gases

The analyzer is suitable for measuring flammable gases in general purpose environment (see page 22).


Oxygen analyzer Magnos28 Measurement principle Paramagnetic behavior of oxygen Magnetomechanical oxygen analyzer Sample component and measurement ranges Sample component Oxygen (O2) Smallest measurement range 0–2 vol.% O2 Measurement range quantity 2 measurement ranges Measurement range limits Measurement ranges are freely adjustable; they are factory-set per order to 0–5 vol.% O2 or 0–25 vol.% O2 or 0–100 vol.% O2 or 98–100 vol.% O2. Largest measurement range 0–100 vol.% O2 Measurement ranges should not be set within ignition limits. Measurement ranges with suppressed zero-point Suppressed measurement ranges are freely adjustable in the range 0–100 vol.% O2. Smallest span 2 vol.% O2. The combination of a suppressed and an initial measurement range is not possible. Pressure correction with a pressure sensor is required. A pressure sensor is installed when the analyzer has been ordered with suppressed measurement range. Stability The following data apply only if all influence factors (e.g. flow rate, temperature, atmospheric pressure) are constant. Linearity deviation ≤ 0.5 % of span, at least 0.005 Vol.-% O2 Repeatability ≤ 50 ppm O2 Zero drift ≤ 3 % of span of the smallest measurement range (per order) per week, minimum 300 ppm O2 per week Sensitivity drift ≤ 0.1 vol.% O2 per week or ≤ 1 % of measured value per week (not cumulative), whichever is smaller; ≤ 0.15 % of measured value per three months, minimum 0.03 vol.% O2 per three months Output fluctuation (2 σ) ≤ 25 ppm O2 at electronic T90 time (static/dynamic) = 3/0 sec Detection limit (4 σ) ≤ 50 ppm O2 at electronic T90 time (static/dynamic) = 3/0 sec Influence effects Flow effect ≤ 0.1 vol.% O2 in the 30–90 l/h range Associated gas effect Data regarding the effect of associated gases can be found in IEC 61207-3:2002 “Gas analyzers – Expression of performance – Part 3: Paramagnetic oxygen analyzers”. Temperature effect Ambient temperature in the permissible range – At zero-point: ≤ 0.02 vol.% O2 per 10 °C – On sensitivity: ≤ 0.3 % vol.% O2 per 10 °C Thermostat temperature = 60 °C Air pressure effect – On sensitivity with no pressure correction: ≤ 1 % of measured value per 1 % air pressure change – On sensitivity with pressure correction using integrated pressure sensor (optional): ≤ 0.1 % of measured value per 1 % air pressure change; for highly suppressed measurement ranges ≤ 0.01 % of measured value per 1 % air pressure change or ≤ 0.002 vol.% O2 per 1 % air pressure change, whichever is greater. Power supply effect DC 24 V ± 5 %: within detection limit Position effect Zero-point shift ≤ 0.05 vol.% O2 per 1° deviation from horizontal orientation. Position has no effect on the hard-mounted unit. 


Dynamic response Warm-up time < 1 hour 90% response time T90 ≤ 3 sec at a sample gas flow of 90 l/h and electronic T90 time (static/dynamic) = 3/0 sec, gas change from nitrogen to air (see “Gas connections”, applies to an analyzer unit only with Magnos28) Calibration Zero-point calibration With oxygen-free process gas or substitute gas End-point calibration With process gas with a known oxygen concentration or a substitute gas such as dried air Single-point calibration Zero-point calibration with any oxygen concentration, e.g. with nitrogen or ambient air, processed through a cooler or H2O absorber. Pressure correction by means of pressure sensor is recommended for single-point calibration with air. Depending on the measurement task involved, the zero- and endpoints should be verified periodically (recommendation: once a year). Calibration of measurement ranges with suppressed zero-point Highly suppressed measurement ranges (≥95–100 vol.% O2) should only be calibrated with test gases with concentrations in the selected measurement range. Single-point calibration can also be done within a suppressed measurement range. The O2 concentration of the test gas must lie within the measurement range. Materials in contact with the sample medium Analyzer Stainless steel 1.4305 (SAE 303), nickel alloy, glass, PtNi, silicon, gold, PTFE; FPM seals, optional: FFKM75 Gas connectors of the solenoid valve (option in model EL3020) PVDF Gas connections See page 31 Sample gas inlet conditions The analyzer must not be used for measurement of ignitable gas/air or gas/oxygen mixtures. Temperature The sample gas dew point should be at least 5 °C below the temperature throughout the sample gas path. Otherwise a sample gas cooler or condensate trap is required. Water vapor content variations cause volume errors. Pressure The analyzer is operated under atmospheric pressure; the sample gas outlet is open to atmosphere. Internal pressure drop < 5 hPa with standard flow rate 60 l/h. Permissible absolute pressure range: 800–1250 hPa. Operation under lower absolute pressure (e.g. at altitudes above 2000 m) on request. Operation under elevated pressure: A pressure sensor is required for pressure influence compensation. Absolute pressure ≤ 1250 hPa: An optional internal pressure sensor can be connected to the sample gas path. Absolute pressure ≥ 1250 hPa: An external pressure sensor must be connected to the sample gas path. Correction is done externally. Flow rate 30–90 l/h Changes of the sample gas flow rate should be avoided in highly suppressed measurement ranges. Corrosive gases Consultation with ABB Analytical is required if the sample gas contains Cl2, HCl, HF or other corrosive components. FFKM75 seals must be used if the sample gas contains NH3. In this case the integral gas feed cannot be connected to the analyzer. The pressure sensor must not be connected to the sample gas path. Flammable gases The analyzer is suitable for measuring flammable gases in general purpose environment (see page 22).


Oxygen analyzer Magnos27 Measurement principle Paramagnetic behavior of oxygen Heavy-duty thermomagnetic analyzer Sample component and measurement ranges Sample component Oxygen (O2) in flue gas or in nitrogen (N2) Smallest measurement range 0–10 vol.% O2 Measurement range quantity 2 measurement ranges Measurement range limits The measurement range is factory-set per customer order. Largest measurement range 0–100 vol.% O2 Measurement ranges within ignition limits cannot be provided. Stability Linearity deviation ≤ 2 % of span Repeatability ≤ 1 % of span Zero drift ≤ 1 % of span per week Sensitivity drift ≤ 2 % of measured value per week Output fluctuation (2 σ) ≤ 0.5 % of smallest measurement range span at electronic T90 time = 0 sec Detection limit (4 σ) ≤ 1 % of smallest measurement range span at electronic T90 time = 0 sec Influence effects Flow effect ≤ 1 % of span at a flow change of ±10 l/h. At an identical flow rate for test and sample gases the flow rate effect is automatically compensated. Associated gas effect Magnos27 calibration applies only to the sample gas shown on the identification plate (= sample component + associated gas). Temperature effect Ambient temperature in permissible range – At zero-point: ≤ 2 % of span per 10 °C – On sensitivity: ≤ 0.5 % of measured value per 10 °C relative to temperature at the time of calibration Thermostat temperature = 63 °C Air pressure effect – At zero-point: < 0.05 vol.% O2 per 1 % air pressure change – On sensitivity without pressure correction: ≤ 1.5 % of measured value per 1 % air pressure change – On sensitivity with pressure correction using integrated pressure sensor (optional): ≤ 0.25 % of measured value per 1 % air pressure change Option: Operating altitude over 2000 m Power supply effect DC 24 V ± 5 %: ≤ 0.2 % of span Position effect Approx. 3 % of smallest measurement range span per 1° deviation from horizontal orientation. Position has no effect on the hardmounted unit.


Dynamic response Warm-up time 2–4 hours 90% response time T90 = 10–22 sec, depending on sample gas flow and on measurement cell connection (see “Gas connections”, applies to an analyzer unit only with Magnos27) Calibration Zero-point calibration With oxygen-free process gas or substitute gas End-point calibration With process gas having a known oxygen concentration or with substitute gas Materials in contact with the sample medium Analyzer Rust- and acid-resistant steel 1.4580 (SAE 316Cb) and 1.4305 (SAE 303), glass Gas lines and connectors Rust- and acid-resistant steel 1.4571 (SAE 316Ti) and 1.4305 (SAE 303), PVC-C, FPM Gas connections See page 32 Sample gas inlet conditions The analyzer must not be used for measurement of flammable gases and ignitable gas/air or gas/oxygen mixtures. Temperature The sample gas dew point should be at least 5 °C below the temperature throughout the sample gas path. Otherwise a sample gas cooler or condensate trap is required. Water vapor content variations cause volume errors. Pressure The analyzer is operated under atmospheric pressure; the sample gas outlet is open to atmosphere. Internal pressure drop < 5 hPa with standard flow rate 60 l/h. Permissible absolute pressure range: 800–1250 hPa. Operation under lower absolute pressure (e.g. at altitudes above 2000 m) on request. Overpressure in the sample cell max. 100 hPa. Flow rate 20–90 l/h Note The analyzer can only be mounted in the 19-inch housing (dimensional drawing see page 32). The analyzer cannot be used in combination with the integral gas feed.


Trace oxygen analyzer ZO23 Measurement principle Potentiometric measurement; zirconium dioxide cell for determination of the oxygen concentration in accordance with Nernst’s equation; reference gas: ambient air. The analyzer is used for the continuous measurement of oxygen in pure gases (N2, CO2, Ar). The measuring cell is catalytically inactivated to the extent that flammable carrier components in stoichiometric concentrations only negligibly reduce the oxygen value. Sample component and measurement ranges Sample component Oxygen (O2) Measurement range quantity 2 measurement ranges Measurement range limits Measurement ranges are freely adjustable within the range 0–1 ppm to 0–250,000 ppm O2; they are factory-set to 0–1/0–10 ppm O2. The following measurement data refer to a measurement span of 100 ppm O2 with a regulated flow rate of 8 ± 0.2 l/h. Stability Linearity Owing to the measurement principle, zirconium dioxide cells are base linear. Repeatability < 1 % of the measurement range or 100 ppb O2 (whichever is greater) Zero drift The zero point (reference point) is displayed if ambient air is present on the sample gas side. The value for air of 20.6 % vol. of O2 (for 25 °C and 50 % relative humidity) may deviate through aging of the cell. < 1 % of the measurement range per week or 250 ppb O2 (whichever is greater) Sensitivity drift Depends on possible interfering components (catalyst poisons) in the sample gas and the aging of the cell. For pure gas measurements in N2 and Ar: < 1 % of the measurement range per week or 250 ppb O2 (whichever is greater) Output fluctuation (2 σ) < ±0.5 % of the measured value or 50 ppb O2 (whichever is greater) Influence effects Flow effect ≤ 300 ppbv O2 in the permissible range Associated gas effect Inert gases (Ar, CO2, N2) have no effect. Flammable gases (CO, H2, CH4) in stoichiometric concentrations to the oxygen content: Conversion of O2 < 20 % of the stoichiometric conversion. If higher concentrations of flammable gases are present, higher O2 conversions must be expected. The concentration of flammable gases in the sample gas must not exceed 100 ppm. Temperature effect The effect of the ambient temperature in the permissible range of +5 to +45 °C is < 2 % of the measured value or 50 ppb O2 per 10 °C change in the ambient temperature (whichever is greater). Air pressure effect No effect through a change in air pressure; the sample gas must flow out of the outlet without back pressure. Power supply effect DC 24 V ± 5 %: no effect Position effect No position effect for permanently installed instruments


Pressure sensor

Use

Standard: with Uras26, Limas23, Caldos27,

Option: with Magnos206, Magnos28 and Magnos27

Working range

pabs = 600–1250 hPa

Materials in contact with the sample medium

Silicone gel, plastics, FPM (Fluorocarbon rubber)

Sample gas composition

The pressure sensor must not be connected to the sample gas

path when the sample gas is corrosive, flammable or explosive.

Fine filtration

Version

Disposable filter with borosilicate glass microfiber filter element

(supplied as accessory)

Retention rate

99.99 % for particles > 0.1 µm

Materials in contact with the sample medium

Polyamide, borosilicate glass with PVDF binder

Safety

Tested per EN 61010-1:2010

Protection class I

Overvoltage category/pollution degree

II/2

Safe isolation

The power supply is galvanically isolated from other circuits by

means of reinforced or double insulation. Protective extra-low

voltage (PELV) on low-voltage side

Electromagnetic compatibility

Noise immunity

Tested to EN 61326-1:2013. Inspection severity: Industrial area,

fulfills at least the test requirements to table 2 of EN 61326.

Emitted interference

Tested to EN 61326-1:2013. Limit value class B for interference

field strength and interference voltage is met.

Mechanical stress

Operation

Vibration test to EN 60068-2-6:1996

In Uras26, slight transient effects on the measured value can occur in the region of the modulation frequency.

Transport

Vibration test to EN 60068-2-6:1996,

shock test to EN 60068-2-27:1995

In its original packaging, the gas analyzer will withstand normal

shipping conditions.

Installation site requirements

Installation location

The gas analyzer is intended for indoor installation only. Installation location altitude max. 2000 m above sea level (over 2000 m

on request)

Ambient temperature

Operation: +5 to +45 °C, Uras26 in combination with another analyzer, Limas23, Fidas24: +5 to +40 °C, Storage and transport:

−25 to +65 °C

Relative humidity

< 75 %, slight condensation allowed

Air circulation

For sufficient air circulation, multiple housings in a 19-inch rack

must be installed with a separation of at least one height unit between housings.

Power supply

Input voltage

AC 100 to 240 V − 15/+ 10 %, 50 to 60 Hz ± 3 Hz

Power consumption

Max. 187 VA

Connection

3-pin plug per EN 60320-1/C14; connection cable supplied.

Connection diagrams see page 24.

Individual analyzers

Input voltage

DC 24 V ± 5 %

Power consumption

Uras26: max. 95 W Magnos27: max. 35 W

Limas23: max. 100 W ZO23: max. 35 W

Magnos206: max. 50 W Caldos27: max. 17 W

Magnos28: max. 50 W Fidas24: max. 40 W

Fidas24: Heating of detector and sample gas inlet

Input voltage

AC 115 V or 230 V, ± 15 % (max. 250 V), 47 to 63 Hz

Power consumption

125 VA for detector heating,

125 VA for heated sample gas inlet (option)

Connection

4-pin plug; connection cable supplied.

Connection diagrams see page 35.

Note regarding the analyzers performance characteristics

The performance characteristics of the analyzers have been determined according to IEC 61207-1:2010 “Expression of performance of gas analyzers – Part 1: General”. They are based on operation at atmospheric pressure (1013 hPa) and nitrogen as the

associated gas. Compliance with these characteristics when

measuring other gas mixtures can only be assured if their composition is known. The physical detection limit is the lower limit

of the performance characteristics relative to the measuring

range span. 











  • ADLINK IMB-M42H ATX Motherboards
  • Adlink PCIe-GIE74 /PCIe-GIE74V Frame Grabbers / Video Capture Cards
  • ADLINK USB-1605 Removable screw terminal blocks or high-density D-Sub
  • Adlink ETX-NR667-L7400/LXE ETX Module
  • ADLINK NuPRO-842LV/P industrial control motherboard
  • ADLINK cPCI-8994 Serial Communication
  • ADLINK / PCI-7432 Digital I/O
  • ADLINK IEC-915GV Compact industrial-grade fanless embedded system
  • ADLINK, PLX / PCI-7296 Digital I/O
  • ADLINK HPCI-14S12U 0040 Backplane Selection Guide
  • ADLINK MXE-5500 /MXE-5501/M16G (G) Fanless Embedded Computer
  • Adlink Multi-Function DAQ PCI-9112
  • ADLINK NuPRO-852LV Control Board
  • ADLINK CompactPCI cPCI-7432R Digital I/O Module
  • ADLINK NUPRO-E320LV PICMG Single Board Computers
  • ADLINK PICMG Single Board Computers NuPRO-E72
  • ADLINK PICMG Single Board Computers NuPRO-865
  • Vibro-Meter Rack VM600SYSMK2-ABE04x 6U 19-inch rack Mk2 system
  • Vibro-Meter Relay expansion RLC16 Mk2 SIL2 (single card) 600-044L
  • Vibro-Meter Relay expansion RLC16 Mk2 SIL2 600-04
  • Vibro-Meter Relay expansion RLC16 Mk2 600-045L
  • Vibro-MeterI/O Relay expansion RLC16 Mk2 600-045
  • Vibro-MeterI/O Relay expansion RLC16 Mk2 600-045
  • Vibro-MeterI/O interface card IOCN Mk2 600-051
  • Vibro-Meter CPU controller   CPUM Mk2   600-050
  • Vibro-Meter I/O interface card    IOC10 Mk2 (rear card)    600-046
  • Vibro-Meter Temperature / Analog Acquisition  AMC10 Mk2  600-042
  • Vibro-Meter I/O interface card IOC4 Mk2 600-043
  • Vibro-Meter Protective measuring card MPC4 Mk2 SIL2 601-041-A10
  • Vibro-Meter Protective measuring card MPC4 Mk2 601-041-A0
  • Vibro-Meter PSU-VM600 VM600 Rack Power Module
  • Vibro-Meter ABE056 - 1U ultra-thin single-rack
  • Vibro-Meter ABE056 - 1U ultra-thin single-rack
  • Vibro-Meter ABE046 - 6U 19-inch rack
  • Vibro-Meter ABE045 - 6U 19-inch rack
  • Vibro-Meter ABE044 - 6U 19-inch rack (with expansion slots)
  • Vibro-Meter ABE042 6U 19-inch rack (4 slots)
  • Vibro-Meter I/O interface card XIO16T
  • Vibro-Meter XMC16 16-channel combustion monitoring card
  • Vibro-Meter XMV16 - 16-channel vibration condition monitoring card
  • Vibro-Meter RLC16 Relay Card
  • Vibro-Meter IRC4 Intelligent Relay Card
  • Vibro-Meter IOCR2 input/output rear interface card
  • Vibro-Meter CPUR2 Central Processing and Controller Card
  • Vibro-Meter I/O interface card  -  IOCR
  • VIBRO-METER CPU controller CPUR
  • VIBRO-METER IOCN 200-585-000-322 I/O interface card
  • VIBRO-METER IOCN 200-585-000-321 I/O Interface Card
  • VIBRO-METER 200-595-074-323 VM600 system CPU management card, non-redundant
  • VIBRO-METER CPUM 200-595-074-322 CPU Controller
  • VIBRO-METER CPU controller CPUM 200-595-070-321
  • Vibro-Meter IOC8T VM600 Mk1 I/O Interface Card
  • Vibro-Meter AMC8 Analog Monitoring Card
  • VIBRO-METER IOC4G2 I/O interface card
  • VIBRO-METER MPC4G2 Protective measuring card
  • VIBRO-METER IOC4T 200-560-101-017 VM600 Mk1 Isolated I/O Interface Card, equipped with MPC4
  • VIBRO-METER IOC4T 200-560-101-015 VM600 Mk1 interface card
  • VIBRO-METER IOC4T 200-560-000-114 I/O interface card
  • VIBRO-METER IOC4T 200-560-000-113 I/O interface card
  • VIBRO-METER IOC4T 200-560-000-111 Sensor wiring
  • VIBRO-METER MPC4 200-510-072-212 Protection Measurement Card
  • VIBRO-METER MPC4 200-510-071-114 Protective Measurement Card
  • VIBRO-METER 200-510-071-113 4-channel vibration protection + 2-speed channels
  • VIBRO-METER 200-510-071-111 4-channel vibration protection + 2-speed channels
  • VIBRO-METER 200-510-070-113 Protective Measurement Card
  • VIBRO-METER 200-510-070-111 4-channel vibration protection + 2-speed channels
  • ADLINK PICMG Single Board Computers NuPRO-E315
  • ADLINK High Speed Link (HSL) PCI-7853
  • ADLINK Digital I/O PCIe-7432
  • ADLINK Mini-ITX Embedded Boards AmITX-AL-I
  • ADLINK 6U CompactPCI 2.0 Blades cPCI-6860A
  • ADLINK PICMG Single Board Computers NuPRO-900A
  • ADLINK Analog Output PCI-6308 Series
  • ADLINK USB DAQ USB-2405
  • ADLINK CRR-LX800-E-500 Compact Industrial Computer Module (COM)
  • ADLINK 15-slot 6U PXI/CompactPCI Chassis with AC PXIS-3320
  • ADLINK NUPRO-842 PICMG Single Board Computers
  • ADLINK PCI-8124-C Advanced 4-CH Encoder Card
  • ADLINK PCI-8102 Centralized Motion Controllers PCI-8102
  • ADLINK PCIe-7200 High-Speed DIO
  • SEL SEL-7201 Feeder Protection Panel
  • SEL SEL-7206 Distribution Bus Differential Protection Panel
  • SEL SEL-4388 Mirrored Bits Tester
  • SEL SEL-T4287 Traveling-Wave Test System
  • SEL SEL-4000 Relay Test System
  • SEL SEL-9510 Control Switch Module
  • SEL SEL-9502 Contact Arc Suppressor
  • SEL SEL-9501 Contact Arc Suppressor
  • SEL SEL-2652 Trip Coil Monitor
  • SEL SEL-2894 Interface Converter
  • SEL SEL-2890 Ethernet Transceiver
  • SEL SEL-2886 EIA-232 to EIA-485 Interface Converter
  • SEL SEL-9220 Fiber-Optic Adapter for SEL-300 Series Relays
  • SEL SEL-2824 Multimode Fiber-Optic EIA-485 Transceivers
  • SEL SEL-2820 Multimode Fiber-Optic EIA-485 Transceivers
  • SEL SEL-2815 Fiber-Optic Transceiver/Modem
  • SEL SEL-2814 Fiber-Optic Transceivers With Hardware Flow Control
  • SEL SEL-2812 Fiber-Optic Transceivers With IRIG-B
  • SEL SEL-2810 Fiber-Optic Transceivers With IRIG-B
  • SEL SEL-2831 Single-Mode Fiber-Optic Transceiver/Modem
  • SEL SEL-2830 Single-Mode Fiber-Optic Transceiver/Modem
  • SEL SEL-2829 Single-Mode Fiber-Optic Transceiver/Modem
  • SEL SEL-C805 200 µm Multimode Fiber-Optic Cableles
  • SEL SEL-C804 Multimode Arc-Flash Detection Fiber-Optic Cables
  • SEL SEL-9331 Power Supply
  • SEL SEL-9322 15 VDC Power Supply
  • SEL SEL-9321 Low-Voltage DC Power Supply
  • SEL SEL-LPS Linear Power Supply
  • SEL SEL-RPM Redundant Power Module
  • SEL SEL-7250 Digital Direct-Replacement Assemblies
  • SEL SEL-2533 Annunciator
  • SEL SEL-2523 Annunciator Panel
  • SEL SEL-2522 Alarm Panel
  • SEL SEL-2516 Rack-Mount Remote I/O Module
  • SEL SEL-2515 Remote I/O Module
  • SEL SEL-2507 High-Speed Remote I/O Module
  • SEL SEL-2506 Rack-Mount Remote I/O Module
  • SEL SEL-2505 Remote I/O Module
  • SEL SEL-3401 Digital Clock
  • SEL SEL-9929 Satellite-Synchronized Clock Display Kit
  • SEL SEL-3405 High-Accuracy IRIG-B Fiber-Optic Transceiver
  • SEL SEL-3400 IRIG-B Distribution Module
  • SEL SEL-2401 Satellite-Synchronized Cloc
  • SEL SEL-2407 Satellite-Synchronized Clock
  • SEL SEL-2488 Satellite-Synchronized Network Clock
  • SEL SEL-5051/5052 Client/Server Network Management System (NMS) Software
  • SEL SEL-5056 Network Management System (NMS)
  • SEL SEL-3031 Serial Radio Transceiverceiver