• HCl– Cation exchanger outlet– Degassed cation conductivity– Semiconductor rinse and reclaim water
• NH3– Ammonia-treated make-up and boiler feed water– Condenser sampling applications– Hot well sampling applications– Before-cation column applications– Inferred pH in ammonia-dosed waters
• Ultra Pure Water (UPW) – Temperature effect of pure water only
• Linear– Manual entry of temperature coefficient
—
‘Trim’ calibration
All ABB sensors are supplied pre-calibrated to exacting
tolerances. This enables installation without the need for any
further adjustment. The construction is robust and provides
highly stable measurements over the sensor’s lifetime.
The single-point calibration facility enables local adjustment
where on-site validation is required
Combined conductivity and pH
The AX416 version measures conductivity and pH with the
same analyzer. The option board provides the capability to
retransmit pH, conductivity and both sample temperatures.
Five programmable alarms can be assigned to suit the needs
of the application, enabling even the most demanding
requirements to be met
Dual input measurement for
critical duties
The AX400 enables two continuous measurements with
simultaneous local display and retransmission. This capability
provides an on-line validation of the process measurement.
Making two simultaneous measurements and configuring
one of the set points as a difference alarm provides
continuous on-line checking and early warning of an invalid
measurement.
AX41x configurable current outputs
The analyzer’s current outputs are fully galvanically isolated
and configurable as 4 to 20 mA, 0 to 20 mA or 0 to 10 mA. In
addition, AX400 conductivity analyzers offer the user a choice
in how the current output is configured:
• Linear
• 2-decade logarithmic
• 3-decade logarithmic
• Bi-linear with configurable break-point
• User-configurable between 0.1 and 100 % of total span
Bi-linear and logarithmic ranges effectively provide a split
output facility. General monitoring is covered with the lower
range while still supplying excellent discrimination during
high level excursions, such as during an alarm state
AX450 and AX455 USP <645> conductivity
AX450 and AX455 analyzers have been developed specifically
with the guidance of major leading pharmaceutical
manufacturers to produce a system fully compliant to
USP<645>.
The US Pharmacopoeia has replaced the off-line qualitative
test of water impurities, such as chloride, with a composite
in-line Stage 1 Test based on conductivity.
Certification
To conform to USP<645> the accuracies of the instrument
display and the sensor’s cell constant are assured below
specified limits. Each AX450 or AX455 analyzer and model
2278 cell is supplied with a traceable test certificate.
Revalidation facility
ABB provides a revalidation facility for the 2278 conductivity
cell where the cell can be returned periodically for re
certification of the cell constant. This quick turnaround
facility fulfills the stringent certification requirements of
today’s pharmaceutical industry.
Recording USP<645>
Two standard current outputs, or four (optional), enable both
conductivity and temperature to be recorded externally as a
permanent record of compliance
Automatic stage 1 test
Maximum acceptable impurity levels are expressed by
USP<645> as uncompensated conductivity at 5 °C (9 °F)
increments. The Stage 1 test of PW and WFI requires both
parameters to be checked against the set limits.
The Stage 1 conductivity / temperature limits are stored
within the AX45X analyzer and can be activated on a limit
alarm contact. The alarm can be configured to initiate at a
safe working distance from the Stage 1 limit. This enables
assurance of operation to Stage 1 without manual
intervention
Display
Type
Dual 5-digit, 7-segment backlit LCD
Information
16-character, single line dot-matrix
Energy-saving function
Backlit LCD configurable as ON or Auto-Off after 60 s
Logbook*
Electronic record of major process events
and calibration data
Real-time clock*
Records time for logbook and auto-manual functions
Relay outputs – on/off
Number of relays
Three supplied as standard or five
with option board fitted
Number of set points
Three supplied as standard or five
with option board fitted
Set point adjustment
Configurable as normal or failsafe high / low
or diagnostic alert
Hysteresis of reading
Programmable 0 to 5 % in 0.1 % increments
Delay
Programmable 0 to 60 s in 1 s intervals
Relay contacts
Single-pole changeover
Rating 5 A, 115 / 230 V AC, 5 A DC
Insulation
2 kV RMS contacts to earth / ground
Analog outputs
Number of current outputs (fully isolated)
Two supplied as standard or four with option board fitted
Output range
• 0 to 10. 0 to 20 or 4 to 20 mA
• Analog output programmable to any value between
0 and 22 mA to indicate system failure
Accuracy
±0.25 % FSD, ±0.5 % of reading (whichever is the greater)
Resolution
0.1 % at 10 mA , 0.05 % at 20 mA
Maximum load resistance
750Ω at 20 mA
Configuration
Can be assigned to either measured variable or
either sample temperature
Digital communications
Communications
Profibus DP (with option board fitted)
Control function – AX410 only
Controller type
P, PI, PID (configurable)
Control outputs