AX410. AX411. AX416. AX450 and AX455
Single and dual input analyzers
for low level conductivity
The AX400 series
The AX400 analyzers incorporate the latest technology to
provide highly reliable, flexible, feature-packed devices that
satisfy a diverse range of process monitoring and control
applications. The complete range encompasses solutions for
pH/Redox (ORP), conductivity and dissolved oxygen.
Models AX41x and AX45x enable continuous measurements
of one or two low-level conductivity points with simultaneous
local display and retransmission. The analyzers are used with
ABB 2-electrode conductivity cells providing measurements
with exceptional accuracy and performance.
Model AX45x analyzers are dedicated to United States
Pharmacopoeia <645>.
AX400 Series analyzers are available for either wall-/pipe- or
panel-mounting and are rated to IP65.
—
High functionality as standard
All versions are supplied with two fully-isolated current
outputs as standard, that can be assigned to the measured
parameter, the sample temperature or any appropriate
calculated variables.
Three programmable relay set points are available that can
also be assigned as required.
Innovative features such as a power saving display and a
diagnostic current output option all contribute to low cost of
ownership
Plug-and-produce expanded control
An advanced function card provides an additional two current
outputs and two further alarm relays that can be assigned to
either measured values or sample temperature.
ABB plug-and-produce software automatically reconfigures
the analyzer if an option board is added later. No user
programming is necessary.
A real-time clock and logbook are also included, making the
full-facility versions extremely powerful and versatile.
Significantly reduced
maintenance costs
The AX400 analyzers are supplied as standard for 85 to
265 V AC operation. There are no inner switches to set.
They can also be provided for 24 V AC or 12 to 30 V DC supply
and recognize automatically which of the two supplies is
being used. 24 V DC operation reduces maintenance costs
significantly by negating the need for costly, yearly safety
tests to ensure compliance with safety procedures.
Energy saving display
The backlit display has been designed to operate in all types
of environments and shows both the measured parameter(s)
and, on a separate 16-character display line, diagnostic and
computed information.
On dual-input analyzers both measured parameters are
displayed simultaneously.
For conservation of energy, the backlight can be set to switch
off automatically after 60 s of inactivity.
Easy access installation terminals
Easy access to the terminations ensures rapid and cost
effective installation. The wall-/pipe-mount version has been
designed to ensure that cable connection is simple and
convenient. Ingress protection of the electronics section is
retained even when the terminal compartment is opened.
Single and dual input analyzers
Both single and dual-input analyzers provide accurate
measurement of low-level conductivity and can be configured
for ranges in µS cm–1. µS m–1. mS cm–1 and mS m–1.
Alternatively, the instrument can be set up for resistivity in
MW-cm, total dissolved solids and salinity.
In addition, the dual-input analyzer calculates, displays and
transmits:
• Ratio of cell 1 / cell 2
• Difference cell 1 – cell 2
• % Passage (cell 1 / cell 2) x 100
• % Rejection (1 – [cell 1 / cell 2]) x 100
• Inferred pH
These facilities are invaluable tools for demineralization and
reverse osmosis and a host of industrial processes.
Easy access installation terminals
Easy access to the terminations ensures rapid and cost
effective installation. The wall-/pipe-mount version has been
designed to ensure that cable connection is simple and
convenient. Ingress protection of the electronics section is
retained even when the terminal compartment is opened.
Cation conductivity and
inferred pH measurement
In low conductivity, ammoniated boiler waters, the AX411
analyzer can calculate an inferred pH measurement from the
conductivity and a preset ammonia concentration. With this
method, one AX411 provides cation conductivity, a direct
output of pH, reduced cost and none of the maintenance
associated with pH electrodes.
Self-monitoring of the validity of the pH measurement is
achieved by checking that an after-cation conductivity value
is sufficiently low. This measurement is provided by the
second input of the AX411 analyzer. Alarm contacts can be
configured for cation conductivity, invalid pH and exhausted
resin
AX410 integral P, PI and PID control
The single input AX410 analyzers incorporate three-term PID
Control offering three modes of sophisticated control:
analog, pulse length (time proportional) and pulse frequency.
These are supplied as standard and can be operated in direct-
or reverse-acting mode, depending upon the application.
Comprehensive temperature
compensations
AX41x and AX45x conductivity analyzers provide true
temperature compensation for all high-purity water
applications:
• Raw (uncompensated) conductivity– USP<645> water for injection and purified water
• NaCl– General monitoring– Mixed-bed exchanger– Final polisher effluent– Cation exchanger inlet– Anion exchanger effluent– Reverse osmosis