DCS Industrial control system
NameDescriptionContent
NEW CENTER
Current Location:

MMS System Manual MMS 6000 Part 2 Direction for use Shaft Vibration Monitor MMS 6110

From:MMS 6110 | Author:LIAO | Time :2025-08-20 | 509 Browse: | 🔊 Click to read aloud ❚❚ | Share:

Please note 

In correspondence concerning this instrument, please quote type number and serial number as given 

on the type plate and software version if applicable. 

Bitte beachten 

Bei Schriftwechsel über dieses Gerät wird gebeten, die Typennummer und die Gerätenummer wie auf 

dem Typenschild aufgedruckt, sowie die Softwareversion (wenn vorhanden), anzugeben 

Important 

As this instrument is an electrical apparatus, it may be operated only by trained personnel. Maintenance and repairs may also be carried out by qualified personnel. 

Wichtig 

Da das Gerät ein elektrisches Betriebsmittel ist, darf die Inbetriebnahme und Bedienung nur durch 

eingewiesenes Personal erfolgen. Wartung und Reparatur dürfen nur von geschultem, fach- und sachkundigen Personal durchgeführt werden. 

PURPOSE AND APPLICATION 

The MMS 6110 dual-channel shaft vibration monitor is part of the MMS 6000 machine monitoring system. In connection with two eddy current measuring chains, the microprocessor 

controlled module serves measurement and supervision of relative shaft vibrations at all kind 

of turbines, compressors, fans and gear boxes. 

During configuration the two channels of the monitor may be defined for separate measurement or for the combined use. The following modes are possible: 

• Dual channel mode with two separate channels.

For each channel the configuration parameters (measuring range, type of transducer, limit 

values etc.), can be defined separately. Each channel calculates its own characteristical 

value (peak-peak or zero-peak value) and checks it on limit exceedings and faults of the 

measuring chain. 

• Calculation and output of maximum value Smax.

The resulting maximum value Smax of the shaft vibration for the measuring directions X 

and Y according to DIN 45670 (characteristical value A) is calculated by geometrical addition of the actual values S1 (channel 1) and S2 (channel 2) [S )t( 1s )t( s2 )t(

2 2

m = + ]. The 

maximum value Smax is output as characteristical value and checked on limit exceedings 

• Calculation and output of maximum vibration amplitudes Sppmax. according to DIN 

45 670, characteristical values B or max (X, Y) according to API 670. 

The characteristical value, output and checked on limit exceedings, is formed by the 

greatest of the two measuring values of channels 1 or 2 in X and Y-measuring direction. 

The characteristical value is output as impressed current of either 0...20 or 4...20 mA. 

Additionally, there is an analog output, providing a standardized d.c. output NGL of 0...+10V 

proportional to the static distance between sensor and measuring target. 

For checking the signals on limit exceedings, there are two alarm channels, each of them 

with one warning (Alert) and one danger level. In the dual channel operating mode the characteristical value is supervised by the relevant alarm channel. The limit values for pre alarm 

(ALERT) and main alarm (DANGER) may be adjusted separately. 

With measuring modes max. value Smax and vibration amplitude Sppmax resp. max(X,Y) the 

characteristical value, calculated from both channels, is supervised by the alarm channels of 

both measuring channels - but with only one "Alert" and one "Danger" limit. 

During configuration the alarm outputs can be activated and the limit values defined. The 

alarm status is shown by two red LEDs on the front of the monitor. The module provides four 

relay driver outputs which can alternatively be switched to normally closed or normally open 

operation mode. 

Input "Alarm blocking" permits blocking of the alarm outputs by means of an external signal. 

During configuration, response delay and alarm blocking at fault situations can be defined. 

Two green LEDs on the front and two relay driver outputs indicate the status of channel and 

monitor supervision. In the normal state, i.e., if neither the channel nor the module supervision have detected an error, if the channel measurement is in the settled state and the 

alarms are not blocked, message "Channel Clear = OK" will be indicated.

If an error occurs and the monitoring function of the module cannot be ensured, this will be 

indicated by flashing or switched off LEDs and by switching off the relevant alarm outputs. 

Configuration of the monitor MMS 6110 is made by means of a laptop computer or a personal computer connected to the RS 232 interface socket on the front of the monitor. The 

software required for configuration and visualization of measuring results and states, as well 

as the connection cable between computer and monitor are part of the MMS 6910 W configuration kit. Moreover, this kit comprises the system manual on the CD-ROM with all necessary information for testing and visualization of measuring results and the states of the 

monitor. The configuration of the monitor is exclusively made by means of the configuration 

software, there are no hardware settings necessary. 

Three different levels permit definition of the access authorization for operation, configuration and for installation and test of the monitor. 

• Access authorization "Factory" includes all adjustments and is intended for installation 

purposes by the epro staff. 

• Access authorization "Service" is intended for specialists who specifies and configures the 

monitor for the actual application. 

• Access authorization „Operation“ permits changing settings required for the normal operation. 

By means of laptop/PC and the operating kit, the parameters of the selected monitor and the 

measuring results (including order analysis, FFT etc.) as well as the status of the monitor can 

be watched on the computer screen. 

The Operating Kit also contains two measuring cables for measuring the sensor signals at 

the mini coax sockets on the monitor front. By means of these cables the coaxial sockets for 

the sensor signals 1 and 2 can be connected to an oscilloscope. 

An RS 485 interface (at the device connector) serves the connection of the monitor to the 

epro MMS 6850 Analysis and Diagnosis system for acquiring and analysing measuring data. 

The monitor is designed as printed board in the standard euro format (100 mm of x 60 mm) 

with 6 TE width (approx. 30 mm) and an eloxadized front plate. 

The supply of the monitor requires +24 V, two supply inputs, decoupled via diodes, permit 

redundant supply of the monitor. 

CONFIGURATION, MENU FILE

This chapter describes the configuration especially of the MMS 6110 module by using the 

MMS 6000W configuration program. The installation software for the configuration program 

MMS 6910 W and the MMS 6000 system manual are part of the MMS 6910 Operating kit 

and stored on a CD-ROM. The description of the handling of the configuration software and 

descriptions of the parameters common to all of the MMS 6000 monitors, are described in 

part 1 - CONFIGURATION AND VISUALIZATION - of the MMS 6000 W System manual. 

There are two different ways how to configure a new monitor: 

1. Select menu "New" in main menu "File", select parameters monitor type and operation mode and set all relevant parameters in menu "Edit". 

2. Call up menu "PreDefined" in menu "File" and select a monitor with an operating 

mode which definitions suits best to the application. At least channel names must be entered and parameters can then be corrected in menu Edit.

Monitor Configuration New

Menupoint File > New opens dialog window Set up MMS-6000 Monitorconfiguration. 

The left part of the menu shows a list where the monitor to be configured is selected. The 

right part of the menu shows a list with all operating modes possible for this monitor, one of 

them must be chosen with a mouse click. 

Possible modes for the MMS 6110: 

S0P and SPP Dual channel mode with two separate channels. The configuration 

(measurement range, sensor type, limit value etc.) can be selected for 

each of the channels, each channel calculates its own characteristical 

value (peak-peak or zero-peak value) and checks it on limit exceedings 

and faults of the measuring chain. 

Smax The resulting maximum value of the shaft vibration for the measuring directions X and Y is calculated according to DIN 45670 characteristical 

value A by geometrical addition of the actual values S1 (channel 1) and 

S2 (channel 2) [S )t( 1s )t( s2 )t(

2 2

m = + ]. The maximum value Smax is output as characteristical value and checked on limit exceedings. 

SPPmax Calculation and output of the greater vibration amplitude Sppmax according to DIN 45670, characteristical value B or max (x, y) according to API 

670. The characteristical value, output and checked on limit exceedings, 

is formed by the greatest of the two measuring values of channels 1 or 2 

in X and Y-measuring direction. 

Having confirmed the choice by clicking the OK- button, the program returns to the main 

menu. The newly defined monitor must then be configured in menu "File > Edit". 

2.2 Monitor type PreDefined 

Menu point File > PreDefined opens dialog window PreDefined MMS 6000 Monitor Types. 

This window shows predefined configurations for different monitor types and their operating 

modes for various applications. After selection of the monitor type and confirmation by clicking the OK button, the configuration is stored in the memory. 

In any case the device parameters must be adapted to the actual application in menu 

File > Edit and the channel designations entered. 

Edit monitor configuration 

Menu point Edit opens a dialog window with several property pages for setting device parameters. 

Property page Administration

In property page Administration the general data, valid for the actual monitor type are 

shown. 

Display field Monitor type

This field shows the predefined monitor type (indicated parameters in this menu cannot be 

changed !) 

Display field Last modification made by

This line shows the names of that user who last made changes of parameters to this monitor 

(cannot be changed in this menu !) 

Display field Last modification on

Date and time of this line indicates the moment when changes to this monitor configuration 

were made last (cannot be changed in this menu !). 

Entry field Factory 

At this point of the menu, the designation for the plant where the machine monitoring system 

is installed, can be entered (max. 41 characters). 

Entry field Block

In this field the designation for the machine block where the monitoring system is installed 

can be typed in (max. 41 characters). 

The entries will be saved with a mouse click on button Apply. 

Property page Basis

Selection of property page Basis opens the menu concerning the device parameters. Access authorization SERVICE permits activation of all functions. Authorization OPERATION 

only permits access to functions limit increase and limit suppression. 

value is calculated by combination of the results of both channels, and watched on limit exceedings, the alarm function will be disabled only, if both channels indicate the message 

"Channel Clear not OK". 

Entry fields Temperature Danger and Temperature Alert 

The input fields for temperature alert alarm and danger alarm show the default factory settings of the alarm values. If the environmental temperature of the monitor exceeds the Alert 

limit measured by a built-in temperature sensor, the Channel Clear LED on the monitor front 

starts flashing quickly. On exceeding the Danger alarm, this is considered to be a module 

fault, as a consequence the Channel Clear LEDs will distinguish and the alarm outputs be 

suppressed (refer to detailed function description -- condition monitoring in section 4.5). 

The predefined limit values of 65 °C and 70 °C may be modified with access authorization 

"Service". However, it has to be taken into account that the danger of measuring faults or 

failures of components increases with higher temperatures. 

Parameter switch Switch on EX-mode

This function has to be activated, if the connected sensors are installed in explosion hazardous areas and must be operated with safety barriers. The activation of this function effects 

that on property page "channel" (channel configuration) the series resistance of the safety 

barriers must be entered into field "series resistance". 

2.3.3 Property page Channel 

The channel configuration is made in property page "Channel". 

  • GE IS200AEPBH1B IS200AEPBH1B‑REV‑AA Pitch-Charger Control PCB Board
  • GE IS200EDEXG1ADA EX2100 SCR Trigger Control Board
  • GE 151X1207CK01SA01 IS200EDEXG1A EX2100 SCR
  • GE 151X1233DD02SA02 IS200AEPBH1B - Variant for pitch control charger assembly
  • GE 151X1233DB01SA01 IS200AEPBH1B - 10kW Wind Power Converter
  • GE 151X1233DB01SA01R | IS200AEPBH1B 10KW Converter
  • GE 151X1233DD01SA02 Pitch Battery Charger (109W6931P001)
  • GE 151X1233DD01SA02 Pitch Battery Charger Assy IS200AEPBH1B
  • GE 151X1233DB02SA02 Rev A IS200AEPBH1B Wind Turbine Converter
  • GE 151X1207CK02SA02 SCR 4400 V 1850 A IS200EDEXG1ADA General Electric EX2100
  • Triconex 3723X 32-Channel TMR Analog Input Module with HART
  • SHINKAWA VM-21G  SIGNAL CONDITIONER SOCKET
  • SHINKAWA VM-21R REVOLUTION SIGNAL CONDITIONER
  • SHINKAWA VM-21T DISPLACEMENT INPUT THRUST SIGNAL CONDITIONER
  • SHINKAWA VM-21D 6-WIRE LVDT SIGNAL CONDITIONER
  • SHINKAWA VM-21P 3-WIRE LVDT SIGNAL CONDITIONER
  • SHINKAWA VM-21A  VIBRATION SIGNAL CONDITIONER
  • SHINKAWA VM-21K DISPLACEMENT INPUT VIBRATION SIGNAL CONDITIONER
  • SHINKAWA VM-25S01 CONDITION MONITORING SYSTEM
  • SHINKAWA VM-25 CONDITION MONITORING SYSTEM
  • SHINKAWA VM-772B DEVICE CONFIG
  • SHINKAWA VM-771B MCL VIEW
  • SHINKAWA VM-76B INSTRUMENT RACK
  • SHINKAWA VM-75B POWER SUPPLY MODULE
  • SHINKAWA VM-742B NETWORK COMMUNICATION MODULE
  • SHINKAWA VM-741B LOCAL COMMUNICATION &  PHASE MARKER MODULE
  • SHINKAWA VM-722B 9-CHANNEL RELAY MODULE
  • SHINKAWA VM-707B AERODERIVATIVE GAS TURBINE  MONITOR MODULE
  • SHINKAWA VM-706B ROD DROP MONITOR MODULE
  • SHINKAWA VM-705B 18-CHANNEL TEMPERATURE   MONITOR MODULE
  • SHINKAWA VM-704B TEMPERATURE MONITOR MODULE
  • SHINKAWA VM-702B ABSOLUTE VIBRATION  MONITOR MODULE
  • SHINKAWA VM-701B VIBRATION / DISPLACEMENT MONITOR MODULE
  • SHINKAWA MODEL VM-5X INTERFACE UNIT
  • SHINKAWA MODEL VM-5W2 DUAL POWER SUPPLY INSTRUMENT RACK
  • SHINKAWA MODEL VM-5H4  INSTRUMENT RACK (for VM-5Z5.6.7)
  • SHINKAWA MODEL VM-5G  SINGLE UNIT   INSTRUMENT RACK
  • SHINKAWA MODEL VM-5Z5,6,7  POWER SUPPLY UNIT
  • SHINKAWA MODEL VM-5Y1,2,3 RELAY MODULE UNIT
  • SHINKAWA MODEL VM-5P3 PHASE MARKER UNIT
  • SHINKAWA MODEL VM-53 DUAL COMMUNICATION UNIT
  • SHINKAWA MODEL VM-52 BOTTOM HOLD MONITOR
  • SHINKAWA MODEL VM-51 ROD DROP MONITOR
  • SHINKAWA MODEL VM-5R TACHOMETER
  • SHINKAWA MODEL VM-5S DUAL TACHOMETER
  • SHINKAWA MODEL VM-5A DUAL VALVE
  • SHINKAWA MODEL VM-5E DUAL CASE
  • SHINKAWA MODEL VM-5L COMPLEMENTARY INPUT
  • SHINKAWA MODEL VM-5N RAMP DIFFERENTIAL
  • SHINKAWA MODEL VM-5D DUAL DIFFERENTIAL
  • SHINKAWA MODEL VM-5T DUAL THRUST MONITOR
  • SHINKAWA MODEL VM-5C ECCENTRICITY MONITOR
  • SHINKAWA MODEL VM-5M DUAL PATH MONITOR
  • SHINKAWA MODEL VM-5B DUAL ACCELERATION MONITOR
  • SHINKAWA MODEL VM-5U DUAL SEISMIC MONITOR
  • SHINKAWA MODEL VM-55 VIBRATION MONITOR
  • SHINKAWA VM-742B1-5 VIBRATION MONITORING SYSTEM CARD
  • Shinkawa PSP-17B P-24-64A P-24-64B Circuit Board Module
  • SHINKAWA MGP-67 PCB
  • Shinkawa JBP-278 P-22-48A-1 P-22-48B-1 Circuit Board Module
  • Shinkawa VM-721B 18-CHANNEL RELAY MODULE
  • Shinkawa FRP-337 P-26-54A P-26-54B Circuit Board Module
  • Shinkawa CPP-111 P-25-59A P-25-59B Circuit Board Module
  • Shinkawa WKN-142K2-2 Vibration Transmitter
  • SHINKAWA VL-202A08L-1012 Rotor Position Monitor Probe Sensor
  • Shinkawa CPP-128 P-26-74A P-26-74B Circuit Board Module
  • Honeywell 51402755-100 PROCESSOR CARD
  • Honeywell 51107595-100 Power
  • HONEYWELL 51304544-100 Control Board
  • HONEYWELL 51109919-100 BOARD
  • HHoneywell 51204160-175 Digital Input Module
  • Honeywell 51304754-150 High Level Analog Input Module
  • HONEYWELL 51304690-100 DIGITAL INPUT CARD
  • Honeywell 51401635-150  COMM CONTROL
  • HONEYWELL 51204174-175 MC-TAOY 52 ANALOG OUTPUT BOARD
  • HONEYWELL  51400596-100
  • HONEYWELL 51401291-100 board card
  • HONEYWELL 51401642-150 I/O LINK MODULE
  • Honeywell LCN‑E Hub Power Supply 38500148‑300
  • Honeywell LCN‑E Hub redundant 38500143‑202 Twisted‑pair LCN
  • Honeywell LCN‑E Hub non‑redundant,38500143‑102
  • Honeywell Enhanced Micro‑TDC‑3000 System Unit MX‑DTAC01K2
  • Honeywell Enhanced Micro‑TDC‑3000 System Unit MX‑DTAB01K2
  • Honeywell Enhanced Micro‑TDC‑3000 System Unit MT11‑520
  • Honeywell STI‑CE Smart Transmitter Interface 51304516‑150
  • Honeywell STI Smart Transmitter Interface 51304516‑100 MC‑PSTX02
  • Honeywell ulse Counter IOP,51304386‑100.MU‑PPIX02
  • Honeywell DO Digital Output IOP‑CE 51304487‑150 CE  version
  • Honeywell DO Digital Output IOP  51304487‑100 MC‑PDOX02
  • Honeywell TDC‑3000 SOE DI IOP,51402625-175 MC‑PDIS12
  • Honeywell TDC‑3000  DI Digital Input IOP‑CE 51304485‑150
  • Honeywell TDC‑3000 DI Digital Input IOP 51304485‑100 MC‑PDIX02
  • Honeywell TDC‑3000 AO Analog Output IOP 51304757‑100 MU‑PAOX02
  • Honeywell TDC‑3000  TC Thermocouple IOP 51304756‑100 MU‑TAMT03
  • Honeywell TDC‑3000   51304755‑100 MU‑TAMR03
  • Honeywell TDC‑3000 LLAI Multiplex AI IOP,51304362‑150
  • Honeywell TDC‑3000  HLAI Analog Input IOP‑CE,51304363‑150
  • Honeywell TDC‑3000  HLAI Analog Input IOP,51304363‑100.4‑20mA
  • Honeywell TDC‑3000  HPM Power Supply,51401731‑100
  • Honeywell TDC‑3000 Standard 1986.HPM High‑Performance PM CPU,51402362‑100
  • Honeywell TDC‑3000 APM Advanced Process Manager CPU,51304459‑100
  • Honeywell TDC‑3000 Standard 1986.PM Process Manager CPU,51304362‑100
  • Honeywell TDC‑3000 NIM Network Interface Module,51304481‑200
  • Honeywell TDC‑3000 NIM Network Interface Module 51304481‑100
  • Honeywell TDC‑3000 BASIC,LCN Terminator Resistor,30732052‑001.Coaxial
  • Honeywell TDC‑3000 BASIC,Hiway Coupler,30753732-001
  • Honeywell TDC‑3000 BASIC,LEPIU Low‑Energy PIU,4DP7APCFA309
  • Honeywell TDC‑3000 BASIC,HLPIU High‑Level PIU,4DP7APCFA106
  • Honeywell TDC‑3000 BASIC,Multifunction Controller(TDC‑2000 DHW),4DP7MFA103
  • Honeywell TDC‑3000 BASIC,Extended Controller(TDC‑2000 DHW),4DP7AXA102
  • Honeywell TDC‑3000 BASIC,Basic Controller(TDC‑2000 DHW),4DP7BBA101
  • Honeywell TDC‑3000 BASIC,XLCNE2 LCN Extender,51305508‑100.Coaxial LCN
  • Honeywell TDC‑3000 BASIC,E‑PLCG PLC Gateway,51400997‑200
  • Honeywell TDC‑3000 BASIC,PLNM Plant Network Module,51401163‑100
  • Honeywell TDC‑3000 BASIC,CG Computer Gateway,80360206‑001
  • Honeywell TDC‑3000 BASIC,HG High‑Speed Gateway,51304526‑100
  • Honeywell TDC‑3000 BASIC,HM History Module,51191596‑200
  • Honeywell TDC‑3000 BASIC,HM History Module,51191596‑100
  • Honeywell TDC‑3000 BASIC,AM Application Module CPU,51401551‑800.K2LCN‑8
  • Honeywell TDC‑3000 BASIC,AM Application Module CPU,51401551‑600.K2LCN‑6
  • Honeywell TDC‑3000 BASIC,AM Application Module CPU,51401551‑400,K2LCN‑4
  • Honeywell TDC‑3000 BASIC,AM Application Module CPU,51401551‑300.K2LCN‑3
  • Honeywell TDC‑3000 BASIC,AM Application Module CPU,51401551‑200,K2LCN‑2
  • Honeywell TDC‑3000 BASIC,US Universal Station CPU,51401288‑200
  • TDC‑3000 BASIC,US Universal Station CPU,51401288‑100
  • SHINKAWA VM-21P1 Signal Conditioner 0-50MM 24VDC
  • Shinkawa RiverNew Vibration Transducer VK202 Series
  • Shinkawa Electric VM-3E2 Vibration Monitor Module