KEY FEATURES AND BENEFITS
• From the vibro-meter ® product line
• Power supply for sensors and/or signal
conditioners with a current output or a
voltage output
• 4 kVRMS galvanic separation between the
sensor side and the monitor side
• 50 VRMS galvanic separation between the
power supply and the output signal
(floating output)
• High rejection of frame voltage
• μA to mV transfer function for current-signal
transmission over longer distances
• V to V transfer function for voltage-signal
transmission over shorter distances
• Available in standard versions and Ex versions
certified for use in potentially explosive
atmospheres (hazardous areas)
• Compatible with all vibro-meter
sensors / measurement chains
• Also compatible with industry-standard IEPE
sensors and 4 to 20 mA loop-powered sensors,
including vibration sensors/transmitters
• No ground connection needed
KEY BENEFITS AND FEATURES (cont’d)
• Suitable for use in functional safety contexts:
SIL 2 in accordance with IEC 61508
• Removable screw-terminal connectors
• DIN-rail mounting
APPLICATIONS
• All vibro-meter sensors / measurement chains
• Safety-related applications
DESCRIPTION
The GSI127 is a galvanic separation unit from
Parker Meggitt’s vibro-meter ® product line. It is
designed for operation with the signal conditioners,
charge amplifiers and electronics (attached or
integrated) used by various vibro-meter
measurement chains and/or sensors.
Compatible devices include the IPC707 signal
conditioners (charge amplifiers) used by CAxxx
piezoelectric accelerometers and CPxxx dynamic
pressure sensors (and older IPC704 signal
conditioners), the IQS9xx signal conditioners used by
TQ9xx proximity sensors (and older IQS4xx signal …
Information contained in this document may be subject to export control regulations of the European Union, USA or other countries. Each
recipient of this document is responsible for ensuring that the transfer or use of any information contained in this document complies with all
relevant export control regulations. ECN N/A.
DESCRIPTION (continued)
conditioners), the attached or integrated electronics
used by CExxx piezoelectric accelerometers, and the
integrated electronics used by the VE210 velocity
sensor. The GSI127 is also compatible with industrystandard IEPE (integrated electronics piezo electric)
sensors such as the CE620, CE630 and PV660
vibration sensors, and 4 to 20 mA loop-powered
sensors such as the CE687 and PV685 vibration
sensors.
Note: The variant of the GSI127 defined by the
ordering option code B11 is compatible with IQS91x
signal conditioners and/or industry-standard
4 to 20 mA loop-powered sensors/transmitters.
The GSI127 galvanic separation unit is a versatile unit
that can be used for the transmission of highfrequency AC signals over long distances in
measurement chains using current-signal
transmission or as a safety barrier unit in
measurement chains using voltage-signal
transmission. More generally, it may be used to
supply any electronic system (sensor side) having a
consumption of up to 21 mA.
The GSI127 also rejects a large amount of the frame
voltage that can introduce noise into a measurement
chain. (Frame voltage is the ground noise and AC
noise pickup that can occur between the sensor case
(sensor ground) and the monitoring system
(electronic ground)). In addition, the design of its
internal power supply results in a floating output
signal, thereby eliminating the need for an additional
external power supply such as a APF19x.
Ex versions of the GSI127 are certified to be installed
in an Ex Zone 2 (“Ex ec”) when supplying
measurement chains installed in Ex environments up
to Zone 0 (“[ ia Ga ]”). These units can also eliminate
the need for additional external Zener barriers.
The GSI127 housing features removable screwterminal connectors that can unplugged from the
main body of the housing to simplify installation and
mounting. It also features a DIN-rail mounting
adaptor for direct mounting on a DIN rail.
In addition, the GSI127 incorporates diagnostics that
automatically and remotely indicates the health/
status of the measurement chain so you always know
when measurements can be trusted. The diagnostics
(SIL 2 “by design”) bring enhanced reliability and
significant risk reduction, making the GSI127 suitable
for use in safety-related applications (functional
safety contexts).
For specific applications, contact your local
Parker Meggitt representative.
BLOCK DIAGRAM

SPECIFICATIONS
Environmental
General
Temperature
• Operating : −20 to 70°C (−4 to 158°F)
• Storage : −40 to 85°C (−40 to 185°F)
Humidity
• Operating : 90% max. non-condensing
• Storage : 95% max. non-condensing
Protection rating
(according to IEC 60529)
: IP20.
Note: The GSI127 is suitable for indoor use only unless it is installed in
an industrial housing or enclosure that ensures a higher level of