are operating correctly. You should replace modules that are not operating
correctly with a spare.
When performed properly, you may remove modules from or install modules into
the rack while power is applied to the rack. Refer to Section 3 for the proper
procedure.
The 3500 Monitoring System is a high precision instrument that requires no
calibration. The functions of the 3500 modules, however, require verification at
regular intervals. You should verify all modules in the 3500 Monitoring System at
these maintenance intervals. The procedures in the Maintenance and
Troubleshooting sections of the module manuals describe the verification and
troubleshooting process.
5.2 Choosing a Maintenance Interval
Use the following approach to choose a maintenance interval:
1. Start with an interval of 1 year and then shorten the interval if any of the
following conditions apply:
• the monitored machine is classified as critical, or
• the 3500 rack is operating in a harsh environment such as in
extreme temperature, high humidity, or a corrosive atmosphere.
2. At each interval, use the results of the previous verifications and ISO
Procedure 10012-1 to adjust the interval.