DCS; Industrial control system
NameDescriptionContent
NEW CENTER
Current Location:

ABBModels EL3020, EL3040

From:ABB | Author:LIAO | Time :2025-08-01 | 197 Browse: | Share:

EasyLine Continuous Gas Analyzers Models EL3020, EL3040 Software Version 3.3 Instructions for Installation, Start-up and Operation Publication No. 41/24400 EN Revision 5 Edition December 2012

Preface

Content of the Operator's Manual

These instructions in brief for installation, start-up and operation contain all the information necessary for the safe and compliant installation, startup and operation of the gas analyzer. Information on calibration, configuration and maintenance of the gas analyzer and the Modbus and Profibus is documented in the on-line help file of the operator's manual; the on-line help file can be found on the CD-ROM "Software tools and technical documentation", which is supplied with the gas analyzer (see below).

Further Information

Analyzer Data Sheet The version of the delivered gas analyzer is described in the "Analyzer Data Sheet" supplied with the gas analyzer. CD-ROM "Software tools and technical documentation" The CD-ROM "Software tools and technical documentation" with the following contents is included in the scope of supply of the gas analyzer:

 Software Tools 

 Operator's Manuals 

 Data Sheets 

 Technical Information 

 Certificates

CD-ROM "Spare parts analytical"

Information on spare parts can be found on the CD-ROM "Spare parts analytical", which is enclosed with the gas analyzer.

Internet

You will find information on ABB Analytical products and services on the Internet at "http://www.abb.com/analytical

Service contact

Service contact If the information in this operator's manual does not cover a particular situation, ABB Service will be pleased to supply additional information as required. Please contact your local service representative. For emergencies, please contact


Basic Steps

The following basic steps should be followed for the installation and commissioning of the gas analyzer:

1 Please note the information on the intended application (see page 9). 

2 Follow safety precautions (see page 10). 

3 Prepare for the installation, provide the requisite material (see page 14). 

4 Unpacking the gas analyzer (see page 41). 

5 Check sample gas path seal integrity (see page 119). 

6 Install the gas analyzer (see page 66). 

7 Connect the gas lines (see page 67). 

8 Connect the electrical leads (see page 79). 

9 Check the installation (see page 90). 

10 Purge the sample gas path (see page 91). 

11 Start up the gas analyzer (see page 92). 

12 Configure the gas analyzer.

Intended Application of the Gas Analyzer

The gas analyzer is designed for continuous measurement of the concentration of individual components in gases or vapors. Any other use is not as specified. The specified use also includes taking note of this operator's manual. The gas analyzer may not be used for the measurement of ignitable gas/air or gas/oxygen mixtures. The gas analyzer version with stainless steel gas lines and gas connections (EL3020 and EL3040 models) may be used for the measurement of flammable gases1 and vapors in a non-hazardous environment, The special requirements for the measurement of flammable gases (see page 39) must be complied with. The oxygen sensor and the modules of the integrated gas feed (Option "Integrated Gas Feed" – only in the EL3020 model, not with Limas23, ZO23, Fidas24) may not be used for the measurement of flammable gases. The explosion-proof version of the gas analyzer with degree of protection II 3G (see page 40) (EL3040 model) may be used for the measurement of non-flammable gases and vapors in a hazardous environment.


NOTE

 The version for the measurement of flammable gases and vapors and the explosion-proof version with degree of protection II 3G are different versions of the gas analyzer and designed for different applications


Safety Information.

Requirements for Safe Operation

In order to operate in a safe and efficient manner the device should be properly handled and stored, correctly installed and set-up, properly operated and correctly maintained.

Personnel Qualifications

Only persons familiar with the installation, set-up, operation and maintenance of comparable devices and certified as being capable of such work should work on the device.

Special Information and Precautions

These include

 The content of this operator's manual, 

 The safety information affixed to the device, 

 The applicable safety precautions for installing and operating electrical devices, 

 Safety precautions for working with gases, acids, condensates, etc.

National Regulations

The regulations, standards and guidelines cited in this operator's manual are applicable in the Federal Republic of Germany. The applicable national regulations should be followed when the device is used in other countries

Device Safety and Safe Operation

The device is designed and tested in accordance with the relevant safety standards and has been shipped ready for safe operation. To maintain this condition and to assure safe operation, read and follow the safety information in this operator's manual. Failure to do so can put persons at risk and can lead to device damage as well as damage to other systems and devices.

Protective Lead Connection

The protective lead (ground) should be attached to the protective lead connector before any other connection is made

Risks of a Disconnected Protective Lead

The device can be hazardous if the protective lead is interrupted inside or outside the device or if the protective lead is disconnected.

Risks Involved in Opening the Covers

Current-bearing components can be exposed when the covers or parts are removed, even if this can be done without tools. Current can be present at some connection points.

Risks Involved in Working with an Open Device

All work on a device that is open and connected to power should only be performed by trained personnel who are familiar with the risks involved.

When Safe Operation can no Longer be Assured

If it is apparent that safe operation is no longer possible, the device should be taken out of operation and secured against unauthorized use. The possibility of safe operation is excluded:

 If the device is visibly damaged, 

 If the device no longer operates, 

 After prolonged storage under adverse conditions, 

 After severe transport stresses.

Measures of the manufacturer

The following measures ensure that the enrichment of combustion gas or an explosive mixture of combustion gas and ambient air cannot occur inside the gas analyzer during normal operation:

 The seal integrity of the combustion gas feed path is checked for a leakage rate of < 1 × 10–4 hPa l/s before delivery. 

 The combustion gas/air mixture (before and after the ignition point) is diluted in the detector with compressed air. 

 The combustion gas feed is not connected to the supply during startup until the internal nominal pressures have been set. 

 The combustion gas feed is switched off if the internal nominal pressures cannot be set during the ignition phase (e.g. because of insufficient compressed air or combustion air feed). 

 The combustion gas feed is switched off after several unsuccessful ignition attempts. 

 If the flame goes out during operation the combustion gas feed is switched off if the following ignition attempts are unsuccessful. The interior of the gas analyzer is not assigned to an (explosion protection) zone; no explosive gas mixture can escape from it to the outside.

Conditions to be complied with by the end user

The end user must comply with the following prerequisites and conditions to ensure safe operation of the gas analyzer:

 The gas analyzer may be used for the measurement of flammable gases provided that the flammable portion does not exceed 15 vol.% CH4 or C1 equivalents. 

 The relevant safety regulations for working with flammable gases must be complied with. 

 The gas connection diagram (see page 64) must be complied with when connecting the combustion gas and combustion air. 

 The combustion gas feed path in the gas analyzer may not be opened! The combustion gas feed path can become leaky as a result! Escaping combustion gas can cause fires and explosions, also outside the gas analyzer! 

 If the combustion gas feed path in the gas analyzer has been opened, however, it must always be checked for seal integrity (see page 118) with a leak detector after it has been sealed again (leakage rate < 1 × 10–4 hPa l/s).

 The seal integrity of the combustion gas line (see page 116) outside the gas analyzer and the combustion gas feed path (see page 118) in the gas analyzer must be checked on a regular basis. 

 The maximum combustion gas flow (see page 30) may not be exceeded. 

 The maximum pressures of combustion gas and combustion air (see page 30) may not be exceeded. 

 The combustion gas flow rate must be limited to a maximum of 10 l/h of H2 or 25 l/h of H2/He mixture. For this purpose, the end user has to provide suitable measures (see page 30) outside the gas analyzer. 

 A shut-off valve (see page 30) must be installed in the combustion gas supply line to increase the safety in the following operating states: 

 Shutting down the gas analyzer, 

 Failure of the instrument air supply, 

 Leakage in the combustion gas feed path inside the gas analyzer. This shut-off valve should be installed outside the analyzer house in the vicinity of the combustion gas supply (cylinder, line). 

 If there is no automatic shut-off of the combustion gas supply to the gas analyzer in case of an instrument supply failure, a visible or audible alarm must be triggered (see page 129). 

 When measuring flammable gases, it must be made sure that in case of a failure of the instrument air supply or of the analyzer module itself the sample gas supply to the analyzer module is shut off and the sample gas path is purged with nitrogen. 

 The unobstructed exchange of air with the environment must be possible around the gas analyzer. The gas analyzer may not be directly covered. The openings in the housing towards the top and at the side may not be closed. The distance to laterally adjacent built-in components must be at least 4 mm. 

 If the gas analyzer is installed in a closed cabinet, adequate ventilation of the cabinet must be provided (at least 1 change of air per hour). The distance to adjacent built-in components towards the top and at the side must be at least 4 mm.




















  • GE SR745-W2-P1-G1-HI-A-L-R-E Feeder protection relay
  • GE IS230TNDSH2A Discrete Output Relay Module Brand
  • GE Fanuc IS200TDBSH2ACC Mark VI Terminal Board Brand
  • GE PMC-0247RC-282000 350-93750247-282000F Disk Drive
  • GE PMC-0247RC-282000 350-93750247-282000F Disk Drive
  • GE VMIVME-1150 Serial Communications Controller
  • GE VMIVME-5576 Fiber-Optic Reflective Memory with Interrupts
  • GE VMIC Isolated Digital Output VMIVME-2170A
  • GE MULTILIN 760 FEEDER MANAGEMENT RELAY 760-P5-G5-S5-HI-A20-R-E
  • GE IS200AEPAH1BKE IS215WEPAH2BB Printed circuit board
  • GE IS210BPPCH1A Mark VIe I/O Pack Processor Card
  • GE IS220PRTDH1A 336A4940CSP6 High-Performance RTD Input Module
  • GE IS220PDIAH1BE 336A5026ADP4 Discrete Input Module
  • GE IS420ESWBH3A IONET Switch Module
  • GE 516TX 336A4940DNP516TX 16-port Ethernet switch
  • GE EVMECNTM13 Embedded control module
  • GE EVPBDP0001 EVPBDP032 control module
  • GE Hydran M2-X Enhanced Monitoring with Extended Sensor Life
  • GE UR6CH Digital I/O Module
  • GE IC695CPU315-CD Central processing unit
  • GE 531X305NTBAMG1 DR Terminal Board
  • GE 531X305NTBALG1 NTB/3TB Terminal Board 531X Series
  • GE 531X305NTBAJG1 NTB/3TB Terminal Board.
  • GE 531X305NTBAHG1 NTB/3TB Terminal Board 531X
  • GE 531X305NTBAEG1 is a PCB that functions as a DR terminal board.
  • General Electric 531X305NTBACG1 NTB/3TB Terminal Board 531X
  • GE Digital Energy D20 Analog Input Module
  • GE 94-164136-001 main board Control board
  • GE 269 PLUS-D/O-100P-125V Digital motor relay
  • GALIL DMC-9940 High-performance motion controller
  • FUJI NP1BS-08 base plate
  • FUJI NP1Y32T09P1 Transistor drain type digital output module
  • FUJI NP1Y16R-08 Digital Output Module
  • FUJI NP1X3206-A High-speed digital input module
  • FUJI NP1AYH4I-MR current output module
  • FUJI NP1S-22 Power module redundancy
  • FUJI RPXD2150-1T servo drive module
  • FUJI FVR008E7S-2UX Ac frequency converter
  • FUJI Ac frequency converter FVR008E7S-2
  • FUJI FVR004G5B-2 Small general-purpose frequency converter
  • FUJI A50L-2001-0232 Industrial control module
  • FUJI A50L-001-0266#N High-performance servo amplifier
  • Honeywell FS7-2173-2RP Gas sensor
  • Honeywell 10106/2/1 Digital Input Module in Stock
  • FRCE SYS68K CPU-40 B/16 PLC core processor module
  • Foxboro FBM I/O cards PBCO-D8-009
  • Foxboro AD916AE Digital Control System (DCS) Module
  • GE SR750-P5-G5-S5-HI-A20-R-E Multilin Relay
  • FOXBORO H90 H90C9AA0117S Industrial Computer Workstation
  • FOXBORO RH928AW | I/A Series Relay Output Module
  • Foxboro N-2AX+DIO Multi-functional input/output module
  • Foxboro RH924WA FCP280 Fiber Optic Network Adapter
  • FOXBORO H92 Versatile Hardware Component In
  • Foxboro FBM218 P0922VW HART® Communication Redundant Output Interface Module
  • Foxboro E69F-TI2-J-R-S E69F Series Current-To-Pneumatic Signal Converter
  • Foxboro E69F-BI2-S Converter
  • Foxboro H92A049E0700 The host of the DCS control station
  • Foxboro H90C9AA0117S Industrial computer workstation
  • Foxboro RH101AA High-performance industrial control module
  • Foxboro P0922YU FPS400-24 I/A Series Power supply
  • FOXBORO P0973LN Chassis-based managed switch with independent power supply
  • FOXBORO P0926PA Input/output module
  • Fanuc A06B-6050-H402 3 AXIS ANALOG AC SERVO DRIVE
  • FOXBORO L0130AD L0130AE-0H Power module group
  • FOXBORO 0399085B 0303440C+0303458A Combination Control Module
  • FOXBORO SY-0399095E (SY-0303451D+SY-0303460E) Process control board
  • FOXBORO 0399071D 0303440C+0303443B Input/Output (I/O) Module
  • FOXBORO RH924UQ Redundant Controller module
  • FFOXBORO E69F-TI2-S current pneumatic converter
  • FOXBORO FBM219 RH916RH Discrete I/O Module
  • FOXBORO FBM227 P0927AC Module
  • FOXBORO 0399144 SY-0301059F SY-1025115C/SY-1025120E I/O module
  • FOXBORO SY-60399001R SY-60301001RB Industrial Control Module
  • FOXBORO 0399143 SY-0301060R SY-1025115C SY-1025120E Combined control board
  • FOXBORO 873EC-JIPFGZ electrodeless conductivity analyzer
  • FOXBORO P0916PH (High-density HART I/O Module)
  • FOXBORO 870ITEC-AYFNZ-7 Intelligent Electrochemical Transmitters
  • FOXBORO Compact FBM240. Redundant with Readback, Discrete
  • FOXBORO FBM208/b, Redundant with Readback, 0 to 20 mA I/O Module
  • FOXBORO FBM201e Analog Input (0 to 20 mA) Interface Modules
  • FOXBORO P0916WG Terminal cable
  • FOXBORO P0926MX 2-Port Splitter
  • FOXBORO AD908JQ High-Frequency Module
  • FOXBORO AD916CC Processor module
  • Foxboro DCS FBM206 Pulse Input Module
  • FOXBORO FBM216 HART® Communication Redundant Input Interface Module
  • Foxboro p0903nu 1×8 unit sub-component module
  • Foxboro P0911SM Industrial control module
  • Foxboro CM902WM I/O module
  • Foxboro CM902WL Power module
  • Foxboro P0972VA Industrial Control Module
  • Foxboro Z-Module Control Processor 270 (ZCP270)
  • Foxboro PO916JS 16-channel terminal block module
  • Foxboro PO911SM High-performance digital/analog input/output module
  • Foxboro P0972PP-NCNI Network Interface Module
  • FOXBORO P0971QZ controller module
  • FOXBORO P0971DP Thermal resistance input/output module
  • FOXBORO P0970VB Cable connector
  • FOXBORO P0970EJ-DNBX Dual-node bus expansion module
  • FOXBORO P0970BP Redundant power supply system
  • FOXBORO P0970BC-DNBI DeviceNet bus interface module
  • FOXBORO P0961FX-CP60S Main control CPU module
  • FOXBORO P0961EF-CP30B Network Interface Unit
  • FOXBORO P0961CA Optical fiber local area network module
  • FOXBORO P0961BD-GW30B gateway processor module
  • FOXBORO P0961BC-CP40B/I/A Series high-performance control processor module
  • FOXBORO P0960JA-CP40 High-performance control processor
  • FOXBORO P0926TM Control module
  • FOXBORO P0916BX Termination Assembly
  • FOXBORO P0916AE P0916AG P0916AW Thermal resistance input type DCS card module
  • FOXBORO P0916AC FOXBORO distributed control system (DCS) compression terminal assembly
  • FOXBORO P0912CB High-performance interface module
  • FOXBORO P0911VJ Thermal resistance input output module
  • FOXBORO P0911QH-A High-precision module
  • FOXBORO P0911QB-C P0911QC-C Thermal resistance input/output module
  • FOXBORO P0904BH P0904FH P0904HB Distributed Control system (DCS) module
  • FOXBORO P0903ZP P0903ZQ Embedded System Debugging Module
  • Foxboro P0903ZL P0903ZN Industrial power module
  • Foxboro P0903ZE I/A Series Fieldbus Isolator Module
  • FOXBORO P0903NW Industrial Control Module
  • FFOXBORO P0903NQ Industrial power module
  • FFOXBORO P0903AA Control Module
  • FOXBORO P0400DL Digital output module
  • FOXBORO P0400BJ Digital output module
  • FOXBORO GW30 industrial control module
  • FOXBORO FBM231 Communication Output Module
  • FOXBORO Fieldbus Communications Module, FCM10Ef
  • FOXBORO Fieldbus Communications Module, FCM10E