DCS; Industrial control system
NameDescriptionContent
NEW CENTER
Current Location:

ABBModels EL3020, EL3040

From:ABB | Author:LIAO | Time :2025-08-01 | 431 Browse: | 🔊 Click to read aloud ❚❚ | Share:

EasyLine Continuous Gas Analyzers Models EL3020, EL3040 Software Version 3.3 Instructions for Installation, Start-up and Operation Publication No. 41/24400 EN Revision 5 Edition December 2012

Preface

Content of the Operator's Manual

These instructions in brief for installation, start-up and operation contain all the information necessary for the safe and compliant installation, startup and operation of the gas analyzer. Information on calibration, configuration and maintenance of the gas analyzer and the Modbus and Profibus is documented in the on-line help file of the operator's manual; the on-line help file can be found on the CD-ROM "Software tools and technical documentation", which is supplied with the gas analyzer (see below).

Further Information

Analyzer Data Sheet The version of the delivered gas analyzer is described in the "Analyzer Data Sheet" supplied with the gas analyzer. CD-ROM "Software tools and technical documentation" The CD-ROM "Software tools and technical documentation" with the following contents is included in the scope of supply of the gas analyzer:

 Software Tools 

 Operator's Manuals 

 Data Sheets 

 Technical Information 

 Certificates

CD-ROM "Spare parts analytical"

Information on spare parts can be found on the CD-ROM "Spare parts analytical", which is enclosed with the gas analyzer.

Internet

You will find information on ABB Analytical products and services on the Internet at "http://www.abb.com/analytical

Service contact

Service contact If the information in this operator's manual does not cover a particular situation, ABB Service will be pleased to supply additional information as required. Please contact your local service representative. For emergencies, please contact


Basic Steps

The following basic steps should be followed for the installation and commissioning of the gas analyzer:

1 Please note the information on the intended application (see page 9). 

2 Follow safety precautions (see page 10). 

3 Prepare for the installation, provide the requisite material (see page 14). 

4 Unpacking the gas analyzer (see page 41). 

5 Check sample gas path seal integrity (see page 119). 

6 Install the gas analyzer (see page 66). 

7 Connect the gas lines (see page 67). 

8 Connect the electrical leads (see page 79). 

9 Check the installation (see page 90). 

10 Purge the sample gas path (see page 91). 

11 Start up the gas analyzer (see page 92). 

12 Configure the gas analyzer.

Intended Application of the Gas Analyzer

The gas analyzer is designed for continuous measurement of the concentration of individual components in gases or vapors. Any other use is not as specified. The specified use also includes taking note of this operator's manual. The gas analyzer may not be used for the measurement of ignitable gas/air or gas/oxygen mixtures. The gas analyzer version with stainless steel gas lines and gas connections (EL3020 and EL3040 models) may be used for the measurement of flammable gases1 and vapors in a non-hazardous environment, The special requirements for the measurement of flammable gases (see page 39) must be complied with. The oxygen sensor and the modules of the integrated gas feed (Option "Integrated Gas Feed" – only in the EL3020 model, not with Limas23, ZO23, Fidas24) may not be used for the measurement of flammable gases. The explosion-proof version of the gas analyzer with degree of protection II 3G (see page 40) (EL3040 model) may be used for the measurement of non-flammable gases and vapors in a hazardous environment.


NOTE

 The version for the measurement of flammable gases and vapors and the explosion-proof version with degree of protection II 3G are different versions of the gas analyzer and designed for different applications


Safety Information.

Requirements for Safe Operation

In order to operate in a safe and efficient manner the device should be properly handled and stored, correctly installed and set-up, properly operated and correctly maintained.

Personnel Qualifications

Only persons familiar with the installation, set-up, operation and maintenance of comparable devices and certified as being capable of such work should work on the device.

Special Information and Precautions

These include

 The content of this operator's manual, 

 The safety information affixed to the device, 

 The applicable safety precautions for installing and operating electrical devices, 

 Safety precautions for working with gases, acids, condensates, etc.

National Regulations

The regulations, standards and guidelines cited in this operator's manual are applicable in the Federal Republic of Germany. The applicable national regulations should be followed when the device is used in other countries

Device Safety and Safe Operation

The device is designed and tested in accordance with the relevant safety standards and has been shipped ready for safe operation. To maintain this condition and to assure safe operation, read and follow the safety information in this operator's manual. Failure to do so can put persons at risk and can lead to device damage as well as damage to other systems and devices.

Protective Lead Connection

The protective lead (ground) should be attached to the protective lead connector before any other connection is made

Risks of a Disconnected Protective Lead

The device can be hazardous if the protective lead is interrupted inside or outside the device or if the protective lead is disconnected.

Risks Involved in Opening the Covers

Current-bearing components can be exposed when the covers or parts are removed, even if this can be done without tools. Current can be present at some connection points.

Risks Involved in Working with an Open Device

All work on a device that is open and connected to power should only be performed by trained personnel who are familiar with the risks involved.

When Safe Operation can no Longer be Assured

If it is apparent that safe operation is no longer possible, the device should be taken out of operation and secured against unauthorized use. The possibility of safe operation is excluded:

 If the device is visibly damaged, 

 If the device no longer operates, 

 After prolonged storage under adverse conditions, 

 After severe transport stresses.

Measures of the manufacturer

The following measures ensure that the enrichment of combustion gas or an explosive mixture of combustion gas and ambient air cannot occur inside the gas analyzer during normal operation:

 The seal integrity of the combustion gas feed path is checked for a leakage rate of < 1 × 10–4 hPa l/s before delivery. 

 The combustion gas/air mixture (before and after the ignition point) is diluted in the detector with compressed air. 

 The combustion gas feed is not connected to the supply during startup until the internal nominal pressures have been set. 

 The combustion gas feed is switched off if the internal nominal pressures cannot be set during the ignition phase (e.g. because of insufficient compressed air or combustion air feed). 

 The combustion gas feed is switched off after several unsuccessful ignition attempts. 

 If the flame goes out during operation the combustion gas feed is switched off if the following ignition attempts are unsuccessful. The interior of the gas analyzer is not assigned to an (explosion protection) zone; no explosive gas mixture can escape from it to the outside.

Conditions to be complied with by the end user

The end user must comply with the following prerequisites and conditions to ensure safe operation of the gas analyzer:

 The gas analyzer may be used for the measurement of flammable gases provided that the flammable portion does not exceed 15 vol.% CH4 or C1 equivalents. 

 The relevant safety regulations for working with flammable gases must be complied with. 

 The gas connection diagram (see page 64) must be complied with when connecting the combustion gas and combustion air. 

 The combustion gas feed path in the gas analyzer may not be opened! The combustion gas feed path can become leaky as a result! Escaping combustion gas can cause fires and explosions, also outside the gas analyzer! 

 If the combustion gas feed path in the gas analyzer has been opened, however, it must always be checked for seal integrity (see page 118) with a leak detector after it has been sealed again (leakage rate < 1 × 10–4 hPa l/s).

 The seal integrity of the combustion gas line (see page 116) outside the gas analyzer and the combustion gas feed path (see page 118) in the gas analyzer must be checked on a regular basis. 

 The maximum combustion gas flow (see page 30) may not be exceeded. 

 The maximum pressures of combustion gas and combustion air (see page 30) may not be exceeded. 

 The combustion gas flow rate must be limited to a maximum of 10 l/h of H2 or 25 l/h of H2/He mixture. For this purpose, the end user has to provide suitable measures (see page 30) outside the gas analyzer. 

 A shut-off valve (see page 30) must be installed in the combustion gas supply line to increase the safety in the following operating states: 

 Shutting down the gas analyzer, 

 Failure of the instrument air supply, 

 Leakage in the combustion gas feed path inside the gas analyzer. This shut-off valve should be installed outside the analyzer house in the vicinity of the combustion gas supply (cylinder, line). 

 If there is no automatic shut-off of the combustion gas supply to the gas analyzer in case of an instrument supply failure, a visible or audible alarm must be triggered (see page 129). 

 When measuring flammable gases, it must be made sure that in case of a failure of the instrument air supply or of the analyzer module itself the sample gas supply to the analyzer module is shut off and the sample gas path is purged with nitrogen. 

 The unobstructed exchange of air with the environment must be possible around the gas analyzer. The gas analyzer may not be directly covered. The openings in the housing towards the top and at the side may not be closed. The distance to laterally adjacent built-in components must be at least 4 mm. 

 If the gas analyzer is installed in a closed cabinet, adequate ventilation of the cabinet must be provided (at least 1 change of air per hour). The distance to adjacent built-in components towards the top and at the side must be at least 4 mm.




















  • ABB 3BUS210755-001 Standard Signal Interface Board
  • ABB 3BUS208800-001 Analog Input/Output Module Card
  • ABB 3BUS208797-001 Standard Signal Condition Board
  • ABB 3BUS208728-001 — Standard Signal Inter Board
  • ABB 3BHT300003R1 CI610 IOB_A Bus Extender for Basic Unit
  • ABB 3BUS208720-001 POWER SIGNAL INTERCONNECT
  • ABB PFVL142R-40MN Ultra-High Capacity Roll Force Load Cell
  • ABB PFVL142R-30MN Industrial Roll Force Load Cell
  • ABBP PFVL142R-25MN Ring-shaped (circular) load cell
  • ABBP FVL142R-20MN Heavy-duty circular load cell
  • ABB PFVL142R-16MN Heavy-duty circular load cell
  • ABB PFVL142R-10MN High-precision and high-strength ring-shaped weighing sensor
  • ABB 3BSE023732R1 PFVO142 Matching unit 25A, 1638 Hz
  • ABB Annular load cells QGPR 104/102
  • ABB Annular load cells PFVL141R
  • ABB Rectangular load cells PFVL141V
  • ABB Circular load cells PFVL141C
  • Metso Automation D201776 ACN PO DC PLC Control Server Computer
  • Nidec MP700A5 / MP700A5R (700A) high-performance DC thyristor drives
  • Nidec Control Techniques MP470A5 / MP470A5R (470A)
  • Nidec Control Techniques MP350A5 / MP350A5R (350A)
  • Nidec Control Techniques MP210A5 / MP210A5R (210A)
  • Nidec Control Techniques MP155A5 / MP155A5R (155A)
  • Nidec Control Techniques MP105A5 / MP105A5R (105A)
  • Nidec Control Techniques MP75A5 / MP75A5R (75A)
  • Nidec Control Techniques MP25A5 / MP25A5R (25A)
  • Nidec Control Techniques MP1850A4 / MP1850A4R (1850A)
  • Nidec Control Techniques MP1200A4 / MP1200A4R (1200A)
  • Nidec Control Techniques MP900A4 / MP900A4R (900A) Overview
  • ABB PFSA240 Roll DC Supply Unit 3BSE073476R1
  • GE AT868-2-1-1 Flow Meter
  • Nidec Mentor MP825A4 / MP825A4R (825A) High-Power DC Drive
  • Nidec Mentor MP700A4 / MP700A4R (700A) DC Drive
  • Nidec MP550A4 / MP550A4R(550A) high-power DC thyristor drive
  • Nidec Mentor MP420A4 / MP420A4R (420A) High-Power DC Drive
  • Nidec Mentor MP350A4 / MP350A4R (350A) High-Power DC Drive
  • Nidec Mentor MP210A4 / MP210A4R(210A) Industrial DC Drive
  • Nidec Mentor MP155A4 / MP155A4R (155A) Industrial DC Drive
  • Nidec Mentor MP105A4 / MP105A4R (105A) Industrial DC Drive
  • Nidec Mentor MP75A4 / MP75A4R (75A) High-Performance DC Drive
  • Nidec MP45A4 / MP45A4R – 45Amp 15kW 4 QUADRANT INDUSTRIAL DCDRIVE
  • Nidec MP25A4 / MP25A4R (25A) Industrial Driver
  • Applied Materials (AMAT) 0100-16009 SMIF-Asyst Wide Body Interlock PCB
  • Applied Materials (AMAT) 0100-09134: DIO Fuse Board Assembly
  • Applied Materials (AMAT) 0100-09107 TEOS Gas Interface Board
  • Applied Materials (AMAT) 0100-09094 TEOS Box Heater Control Board
  • Applied Materials (AMAT) 0100-09071 SBC I/O Breakout Board
  • Applied Materials (AMAT) 0100-09010 System Electronics Backplane
  • Applied Materials (AMAT) 0100-09006 Intelligent Interface SBC Module
  • Applied Materials (AMAT) 0100-03920 Analog I/O Interface Board
  • Applied Materials (AMAT) 0100-00511 / 0130-00511: Mainframe I/O Display
  • Applied Materials (AMAT) 0100-00156 Precision Isolation Amplifier Board
  • Applied Materials (AMAT) 0100-00052 High-Performance Communication Interface
  • Autronica Controller rack unit BC-250
  • Autronica Mimic cabinet BUR-200. for maritime installations
  • Autronica Mimic driver BUR-200
  • Autronica Information panel BV-210
  • Autronica Fire brigade panel BU-210
  • Autronica Repeater panel BS-211
  • Autronica Maritime fire alarm control panel BS-200M
  • Autronica Fire alarm control panel BS-200
  • AMAT 0100-00046 AC Current Sense PWB
  • AMAT A0414720 Precision Advanced System Controller
  • AMAT 0010-00017 Precision Semiconductor Process Interface
  • AMAT 01-82889-00 High-Performance Semiconductor Component
  • ABB Sample Gas Cooler SCC-C 23070-0-10232110
  • IBA ibaRackline-PCHD Efficient process analysis with ibaHD-Server
  • IBA ibaRackline-PC CAM Frame-accurate video information with ibaCapture
  • IBA ibaRackline-PC Highly available and reliable
  • IBA Optical Signal Multiplier ibaBM-FOX-i-3o-D
  • IBA Optical Data Distribution System ibaBM-DIS-i-8o
  • IBA Optical Data Concentrator ibaBM-COL-8i-o
  • IBA ibaNet750-BM-D Acquisition via FO
  • IBA ibaW-750 Acquisition via Ethernet
  • IBA ibaPADU-8AI-I Compact Measurement Modules
  • IBA ibaPADU-D-8AI-I Compact Measurement Modules
  • IBA ibaPADU-8AI-U Compact Measurement Modules
  • IBA ibaPADU-D-8AI-U Compact Measurement Modules
  • IBA ibaPADU-4-AI-U Compact Measurement Modules
  • IBA ibaPADU-C-8AI Self-Supplied Data Logger
  • IBA ibaBM-ENetIP Bus monitor for EtherNet/IP
  • IBA ibaBM-eCAT Bus monitor for EtherCAT
  • IBA ibaBM-DP Bus monitor for PROFIBUS
  • IBA ibaBM-PN: Bus monitor for PROFINET IO
  • IBA ibaMS3xAI-1A Precision AC Current Measurement Module
  • IBA ibaDAQ Intelligent Central Unit
  • IBA ibaPQU-S Power Quality Monitoring System
  • IBA ibaCMU-S Condition Monitoring Unit (CMU)
  • IBA ibaPADU-S-IT-2x16 Modular data acquisition and control system
  • IBA ibaPADU-S-CM Modular data acquisition system
  • IBA ibaM-4AI-IEPE Input module
  • IBA ibaM-4AI-UI Input module
  • IBA ibaM-4AI-150V-AC Input module
  • IBA ibaM-4AI-600V-AC Input module
  • IBA ibaLink-SM-256V: High-Density PLC Data Interface
  • IBA ibaLink-SM-64V High-Performance S5/S7 Interface
  • IBA ibaLink-SM-128V-i-2o Synchronous Fiber Optic (ibaNet)
  • IBA ibaLink-SM-128V communication module
  • IBA ibaM-4AI-5A-150A-AC Input module
  • IBA ibaM-FO-2IO Interface module
  • IBA ibaM-COM Communication module
  • IBA ibaM-DAQ Intelligent Processor Module
  • B&R ECE161-0 MULTI digital input module
  • B&R ECCP70-01 MULTI CPU type B 42 KByte SRAM
  • B&R ECCP60-01 MULTI CPU type B 42 KByte SRAM
  • B&R DI426 digital input module
  • B&R 2DS100.60-1 electronic drum sequencer Absolut encoder
  • B&R 2CP100.60-1 CPU MODULE
  • B&R 2BM100.9 High-performance I/O module
  • AMAT 0190-14928 SCR Power Controller (PVD Reverse Zone)
  • AMAT 0500-01065 300mm Loadlock Interface Interlock Board
  • AMAT 2000-21123 Advanced Vacuum Seal Assembly
  • AMAT 0660-00090 High-Performance Industrial Power Filter
  • AMAT 0240-34077 Centura Endpoint Controller Kit
  • AMAT 0195-10215 High-Precision Pedestal Assembly
  • AMAT 0190-76050 VGA Video Controller VME Module
  • AMAT 0190-75084 High-Performance Communication & Logic Controller
  • AMAT 0190-60287 Precision VME/cPCI Interface Control Module
  • AMAT 0190-53752 DI Water I/O Controller PCB
  • AMAT 0190-37993 DeviceNet Scanner Pro (3U CompactPCI)
  • AMAT 0190-37833 MKS CDN500R-5 EPI 300mm Interface Module
  • AMAT 0190-37771 MKS CDN500R Interlock Control Module
  • AMAT 0190-37616 High-Precision Analog Input/Output Interface
  • AMAT 0190-36787B ISAC CP I/O Block 2 (Top) - Revision B
  • AMAT 0190-36787 ISAC CP I/O Block 2 (Top)
  • AMAT 0190-36511 DIP294 DeviceNet I/O Control Block
  • AMAT 0190-35764 & 0190-35765: Precision Control Interface Duo
  • AMAT 0190-35763 High-Performance Integrated Power Module