DCS Industrial control system
NameDescriptionContent
NEW CENTER
Current Location:

ABBModels EL3020, EL3040

From:ABB | Author:LIAO | Time :2025-08-01 | 636 Browse: | 🔊 Click to read aloud ❚❚ ▶ | Share:

EasyLine Continuous Gas Analyzers Models EL3020, EL3040 Software Version 3.3 Instructions for Installation, Start-up and Operation Publication No. 41/24400 EN Revision 5 Edition December 2012

Preface

Content of the Operator's Manual

These instructions in brief for installation, start-up and operation contain all the information necessary for the safe and compliant installation, startup and operation of the gas analyzer. Information on calibration, configuration and maintenance of the gas analyzer and the Modbus and Profibus is documented in the on-line help file of the operator's manual; the on-line help file can be found on the CD-ROM "Software tools and technical documentation", which is supplied with the gas analyzer (see below).

Further Information

Analyzer Data Sheet The version of the delivered gas analyzer is described in the "Analyzer Data Sheet" supplied with the gas analyzer. CD-ROM "Software tools and technical documentation" The CD-ROM "Software tools and technical documentation" with the following contents is included in the scope of supply of the gas analyzer:

 Software Tools 

 Operator's Manuals 

 Data Sheets 

 Technical Information 

 Certificates

CD-ROM "Spare parts analytical"

Information on spare parts can be found on the CD-ROM "Spare parts analytical", which is enclosed with the gas analyzer.

Internet

You will find information on ABB Analytical products and services on the Internet at "http://www.abb.com/analytical

Service contact

Service contact If the information in this operator's manual does not cover a particular situation, ABB Service will be pleased to supply additional information as required. Please contact your local service representative. For emergencies, please contact


Basic Steps

The following basic steps should be followed for the installation and commissioning of the gas analyzer:

1 Please note the information on the intended application (see page 9). 

2 Follow safety precautions (see page 10). 

3 Prepare for the installation, provide the requisite material (see page 14). 

4 Unpacking the gas analyzer (see page 41). 

5 Check sample gas path seal integrity (see page 119). 

6 Install the gas analyzer (see page 66). 

7 Connect the gas lines (see page 67). 

8 Connect the electrical leads (see page 79). 

9 Check the installation (see page 90). 

10 Purge the sample gas path (see page 91). 

11 Start up the gas analyzer (see page 92). 

12 Configure the gas analyzer.

Intended Application of the Gas Analyzer

The gas analyzer is designed for continuous measurement of the concentration of individual components in gases or vapors. Any other use is not as specified. The specified use also includes taking note of this operator's manual. The gas analyzer may not be used for the measurement of ignitable gas/air or gas/oxygen mixtures. The gas analyzer version with stainless steel gas lines and gas connections (EL3020 and EL3040 models) may be used for the measurement of flammable gases1 and vapors in a non-hazardous environment, The special requirements for the measurement of flammable gases (see page 39) must be complied with. The oxygen sensor and the modules of the integrated gas feed (Option "Integrated Gas Feed" – only in the EL3020 model, not with Limas23, ZO23, Fidas24) may not be used for the measurement of flammable gases. The explosion-proof version of the gas analyzer with degree of protection II 3G (see page 40) (EL3040 model) may be used for the measurement of non-flammable gases and vapors in a hazardous environment.


NOTE

 The version for the measurement of flammable gases and vapors and the explosion-proof version with degree of protection II 3G are different versions of the gas analyzer and designed for different applications


Safety Information.

Requirements for Safe Operation

In order to operate in a safe and efficient manner the device should be properly handled and stored, correctly installed and set-up, properly operated and correctly maintained.

Personnel Qualifications

Only persons familiar with the installation, set-up, operation and maintenance of comparable devices and certified as being capable of such work should work on the device.

Special Information and Precautions

These include

 The content of this operator's manual, 

 The safety information affixed to the device, 

 The applicable safety precautions for installing and operating electrical devices, 

 Safety precautions for working with gases, acids, condensates, etc.

National Regulations

The regulations, standards and guidelines cited in this operator's manual are applicable in the Federal Republic of Germany. The applicable national regulations should be followed when the device is used in other countries

Device Safety and Safe Operation

The device is designed and tested in accordance with the relevant safety standards and has been shipped ready for safe operation. To maintain this condition and to assure safe operation, read and follow the safety information in this operator's manual. Failure to do so can put persons at risk and can lead to device damage as well as damage to other systems and devices.

Protective Lead Connection

The protective lead (ground) should be attached to the protective lead connector before any other connection is made

Risks of a Disconnected Protective Lead

The device can be hazardous if the protective lead is interrupted inside or outside the device or if the protective lead is disconnected.

Risks Involved in Opening the Covers

Current-bearing components can be exposed when the covers or parts are removed, even if this can be done without tools. Current can be present at some connection points.

Risks Involved in Working with an Open Device

All work on a device that is open and connected to power should only be performed by trained personnel who are familiar with the risks involved.

When Safe Operation can no Longer be Assured

If it is apparent that safe operation is no longer possible, the device should be taken out of operation and secured against unauthorized use. The possibility of safe operation is excluded:

 If the device is visibly damaged, 

 If the device no longer operates, 

 After prolonged storage under adverse conditions, 

 After severe transport stresses.

Measures of the manufacturer

The following measures ensure that the enrichment of combustion gas or an explosive mixture of combustion gas and ambient air cannot occur inside the gas analyzer during normal operation:

 The seal integrity of the combustion gas feed path is checked for a leakage rate of < 1 × 10–4 hPa l/s before delivery. 

 The combustion gas/air mixture (before and after the ignition point) is diluted in the detector with compressed air. 

 The combustion gas feed is not connected to the supply during startup until the internal nominal pressures have been set. 

 The combustion gas feed is switched off if the internal nominal pressures cannot be set during the ignition phase (e.g. because of insufficient compressed air or combustion air feed). 

 The combustion gas feed is switched off after several unsuccessful ignition attempts. 

 If the flame goes out during operation the combustion gas feed is switched off if the following ignition attempts are unsuccessful. The interior of the gas analyzer is not assigned to an (explosion protection) zone; no explosive gas mixture can escape from it to the outside.

Conditions to be complied with by the end user

The end user must comply with the following prerequisites and conditions to ensure safe operation of the gas analyzer:

 The gas analyzer may be used for the measurement of flammable gases provided that the flammable portion does not exceed 15 vol.% CH4 or C1 equivalents. 

 The relevant safety regulations for working with flammable gases must be complied with. 

 The gas connection diagram (see page 64) must be complied with when connecting the combustion gas and combustion air. 

 The combustion gas feed path in the gas analyzer may not be opened! The combustion gas feed path can become leaky as a result! Escaping combustion gas can cause fires and explosions, also outside the gas analyzer! 

 If the combustion gas feed path in the gas analyzer has been opened, however, it must always be checked for seal integrity (see page 118) with a leak detector after it has been sealed again (leakage rate < 1 × 10–4 hPa l/s).

 The seal integrity of the combustion gas line (see page 116) outside the gas analyzer and the combustion gas feed path (see page 118) in the gas analyzer must be checked on a regular basis. 

 The maximum combustion gas flow (see page 30) may not be exceeded. 

 The maximum pressures of combustion gas and combustion air (see page 30) may not be exceeded. 

 The combustion gas flow rate must be limited to a maximum of 10 l/h of H2 or 25 l/h of H2/He mixture. For this purpose, the end user has to provide suitable measures (see page 30) outside the gas analyzer. 

 A shut-off valve (see page 30) must be installed in the combustion gas supply line to increase the safety in the following operating states: 

 Shutting down the gas analyzer, 

 Failure of the instrument air supply, 

 Leakage in the combustion gas feed path inside the gas analyzer. This shut-off valve should be installed outside the analyzer house in the vicinity of the combustion gas supply (cylinder, line). 

 If there is no automatic shut-off of the combustion gas supply to the gas analyzer in case of an instrument supply failure, a visible or audible alarm must be triggered (see page 129). 

 When measuring flammable gases, it must be made sure that in case of a failure of the instrument air supply or of the analyzer module itself the sample gas supply to the analyzer module is shut off and the sample gas path is purged with nitrogen. 

 The unobstructed exchange of air with the environment must be possible around the gas analyzer. The gas analyzer may not be directly covered. The openings in the housing towards the top and at the side may not be closed. The distance to laterally adjacent built-in components must be at least 4 mm. 

 If the gas analyzer is installed in a closed cabinet, adequate ventilation of the cabinet must be provided (at least 1 change of air per hour). The distance to adjacent built-in components towards the top and at the side must be at least 4 mm.




















  • AEROTECH SM Series High-Performance Stepper Motors
  • AEROTECH BLMSC Series Linear Motors
  • AEROTECH Custom Linear Motors
  • AEROTECH BMS Series Brushless, Slotless Rotary Motors
  • AEROTECH BM Series Brushless Rotary Motors
  • AEROTECH BLMX Series Linear Motors
  • AEROTECH BLMH Series Linear Motors
  • AEROTECH BLM Series Linear Motors
  • AEROTECH BLMC Series Linear Motors
  • AEROTECH BLMUC-143  Linear motor forcer with thermistor
  • AEROTECH BLMUC-111    Linear motor forcer with thermistor
  • AEROTECH BLMUC-95  Linear motor forcer with thermistor
  • AEROTECH BLMUC-79 Linear motor forcer with thermistor
  • AEROTECH BLMUC Series Linear Motors
  • AEROTECH Automation1 FLEX Distributed Amplifier Multi-Axis Servo Control System
  • AEROTECH Automation1 HXA4 6-Axis Hexapod Servo Drive
  • AEROTECH Automation1 iHXA4 6-Axis Hexapod Servo Drive with Motion Controller
  • AEROTECH Modular System Control Console
  • AEROTECH TV Series Transformers
  • AEROTECH TM Series Transformers
  • AEROTECH UFM Noise Filter Module
  • AEROTECH MFM Motor Filter Module
  • AEROTECH MPG Handwheel Pendant
  • AEROTECH Automation1 Multi-Axis Joystick
  • AEROTECH ETM Enhanced Throughput Module
  • AEROTECH Automation1 XL4s High-Speed, Single-Phase Linear Servo Drive
  • AEROTECH Automation1 XL2e Enhanced, Compact Linear Servo Drive
  • Automation1 iXL2e Enhanced, Compact Linear Servo Drive with Motion Controller
  • AEROTECH Automation1 XL5e Enhanced, Linear Servo Drive
  • AEROTECH Automation1 iXL5e Enhanced Linear Servo Drive with Motion Controller
  • AEROTECH Automation1 GI4 Laser Scan Head Controller
  • AEROTECH Automation1 SI4 Stepper Interface
  • AEROTECH Automation1 XI4 Servo Interface
  • AEROTECH AEROTECH Automation1 iXI4 Servo Interface with Motion Controller
  • AEROTECH Automation1 XC6e Enhanced, High-Power PWM Servo Drive
  • AEROTECH Automation1 iXC6e Enhanced, High-Power PWM Servo Drive with Motion Controller
  • AEROTECH Automation1 XC2 Compact PWM Servo Drive
  • AEROTECH Automation1 XC4 PWM Servo Drive
  • AEROTECH Automation1 GL4 Galvo Laser Scan Head Linear Drive
  • AEROTECH Automation1 XC2e Enhanced, Compact Servo Drive with PWM+
  • AEROTECH Automation1 XC4e Enhanced, PWM Servo Drive
  • AEROTECH Automation1 XR3 Multi-Axis Servo Drive Rack
  • AEROTECH Automation1 iXR3 Multi-Axis Servo Drive Rack with Motion Controller
  • AEROTECH Automation1 iXC2 Compact PWM Servo Drive with Motion Controller
  • AEROTECH Automation1 iXC2e Enhanced, Compact PWM+ Servo Drive with Motion Controller
  • AEROTECH Automation1 iXC4 PWM Servo Drive with Motion Controller
  • AEROTECH Automation1 iXC4e Enhanced, PWM Servo Drive with Motion Controller
  • AEROTECH Automation1 iPC Industrial PC with Motion Controller
  • AEROTECH EPICS & TANGO Drivers for Distributed Control Systems
  • AEROTECH .NET, C, Python & LabVIEW APIs Custom Application Develotpmen
  • AEROTECH Automation1 Industrial Ethernet Support
  • AEROTECH Automation1 iSMC Intelligent, Software-Based Motion Controller
  • AEROTECH Automation1 iXA4 1- 2- or 4-Axis PWM Servo Drive with Motion Controller
  • AEROTECH Automation1 XA4 1- 2- or 4-Axis PWM Servo Drive
  • CTI 2500-ADP2-DISC Discrete Wiring Adapter for Series 500
  • CTI 2500-ADP1 Analog Wiring Adapter for Series 500
  • CTI 2500C-RADP-RS485 Remote Base Controller Adapter for Series 500
  • CTI 2500C-RADP-RBC Profibus RBC Adapter for Series 500
  • CTI 2500C-PADP-120V 120VAC 50W Power Supply for Series 500
  • CTI 2500-RADP RBC Adapter for Series 500
  • CTI 2500-R4500 520/530/DBC Adapter for Series 500 - MATURE
  • CTI 2500-PADP Power Supply Adapter for Series 500
  • CTI 2500-IADP I/O Adapter for Series 500
  • CTI 2500-VP15A-N4-W7 15" Flat-Panel PC with Windows® 7 Embedded
  • CTI 2500S-27-1750 Slice I/O, Ethernet, RS232/485. 900MHz, 2DIO / 4UAI / 2AI
  • CTI 2500S-26-1750 Slice I/O, Ethernet, RS232/485. 900MHz, 8DIO / 2UAI
  • CTI 2500S-23-1750 Slice I/O, Ethernet, RS232/485. 900MHz, 4DI / 4DO / 2UAI / 2AO
  • CTI 2500S-27-1550 Slice I/O, RS232/485. 2DIO / 4UAI / 2AI
  • CTI 2500S-26-1550 Slice I/O, RS232/485. 8DIO / 2UAI
  • CTI 2500S-23-1550 Slice I/O, RS232/485. 4DI / 4DO / 2UAI / 2AO
  • CTI 2500S-27-1050 Slice I/O, Ethernet, 2DIO / 4UAI / 2AI
  • CTI 2500S-26-1050 Slice I/O, Ethernet, 8DIO / 2UAI
  • CTI 2500S-23-1050 Slice I/O, Ethernet, 4DI / 4DO / 2UAI / 2AO
  • CTI 2500C-32F-CJC 32-Pin Field Wiring Connector with Cold Junction Compensation
  • CTI 2500C-32F 32-Pin Field Wiring Connector
  • CTI 2500C-8-TC Eight Thermocouple Inputs
  • CTI 2500C-8-RTD Eight RTD Inputs
  • CTI 2500C-8-RL-FC 8 Form-C Relay Outputs
  • CTI 2500C-8-IDO-24V 8 Isolated 24VDC Outputs
  • CTI 2500C-8-IDO-120V 8 Isolated 120VAC Outputs
  • CTI 2500C-16-DO-24V Sixteen 24VDC Digital Outputs
  • CTI 2500C-16-DO-120V Sixteen 120VAC Digital Outputs
  • CTI 500C-16-IDI-24V Sixteen Isolated 24V AC/DC Digital Inputs
  • CTI 2500C-16-IDI-120V Sixteen Isolated 120V AC/DC Digital Inputs
  • CTI 2500C-16-DIDO-SIM Sixteen Digital Input and Output SImulator
  • CTI 2500C-16-DI-24V Sixteen 24V AC/DC Digital Inputs
  • CTI 2500C-16-DI-120V Sixteen 120V AC/DC Digital Inputs
  • CTI 2500C-RBC-RS485 RS485 Remote Base Controller
  • CTI 2500C-RBC-PRF Profibus Remote Base Controller
  • CTI 2500C-PS-24V-35 35-Watt DC Power Supply
  • CTI 2500C-PS-120V-35 35-Watt AC Power Supply
  • CTI 2500C-R8 Eight-Slot Base
  • CTI 2500C-R4 Four-Slot Base
  • CTI 2500C-R16 Sixteen-Slot Base
  • CTI 2500C-2572-B Fast Ethernet TCP/IP Adapter
  • CTI 2500C-J750 Compact Janus PAC with 3MB Project Memory
  • CTI 2500C-C300 CPU with 512K User Memory
  • CTI 2500C-C200 CPU with 256K User Memory
  • CTI 2500C-C100 CPU with 128K User Memory
  • CTI 2500C-8-AO 8 Analog Outputs
  • CTI 2500C-8-AI 8 Analog Inputs
  • CTI 2559-FCAL Precision Calibration Connector
  • CTI 2559-FPC 40-Position Screw-Terminal Connector
  • CTI 2500-RFC PLC to RF Modem Cable
  • CTI 2500-40F 40-Position Standard Screw-Terminal Connector
  • CTI 2505 Vibration Sensor Interface Module
  • CTI 2554-A 4-Channel Isolated High Speed Counter Module
  • CTI 2553-A 2-Channel Mag Meter Input Module
  • CTI 2502 2-Channel High Speed Counter Input Module
  • CTI 2559-TC 8-Channel Thermocouple Input Module
  • CTI 2556-A 16-Channel Isolated Thermocouple Input Module
  • CTI 2556 16-Channel Isolated Thermocouple Input Module
  • CTI 2551-A 8-Channel Isolated Thermocouple Input Module
  • CTI 2559-RTD 8-Channel RTD Input Module
  • CTI 2557-A 16-Channel RTD Input Module
  • CTI 2557 16-Channel RTD Input Module
  • CTI 2552-A 8-Channel Isolated RTD Input Module
  • CTI 2534 8-Point Form-C Relay Output Module
  • CTI 2532 16-Point Form-A Relay Output Module
  • CTI  2531 32-Point Form-A Relay Output Module
  • CTI  2530 8-Point Form-C Relay Output Module
  • CTI  2562 8-Channel Analog Output Module
  • CTI  2560-A 8-Channel Isolated Analog Output Module
  • CTI  2501 8in/4out Analog Module
  • CTI  2558 8-Channel Analog Input Module
  • CTI  2555-A 16-Channel Analog Input Module
  • CTI  2550-A 8-Channel Isolated Analog Input Module
  • CTI  2599 8/16/32-Point AC Output Module