Pressure sensor
Use
Standard: with Uras26, Limas23, Caldos27,
Option: with Magnos206, Magnos28 and Magnos27
Working range
pabs = 600–1250 hPa
Materials in contact with the sample medium
Silicone gel, plastics, FPM (Fluorocarbon rubber)
Sample gas composition
The pressure sensor must not be connected to the sample gas
path when the sample gas is corrosive, flammable or explosive.
Fine filtration
Version
Disposable filter with borosilicate glass microfiber filter element
(supplied as accessory)
Retention rate
99.99 % for particles > 0.1 µm
Materials in contact with the sample medium
Polyamide, borosilicate glass with PVDF binder
Safety
Tested per EN 61010-1:2010
Protection class I
Overvoltage category/pollution degree
II/2
Safe isolation
The power supply is galvanically isolated from other circuits by
means of reinforced or double insulation. Protective extra-low
voltage (PELV) on low-voltage side
Electromagnetic compatibility
Noise immunity
Tested to EN 61326-1:2013. Inspection severity: Industrial area,
fulfills at least the test requirements to table 2 of EN 61326.
Emitted interference
Tested to EN 61326-1:2013. Limit value class B for interference
field strength and interference voltage is met.
Mechanical stress
Operation
Vibration test to EN 60068-2-6:1996
In Uras26, slight transient effects on the measured value can occur in the region of the modulation frequency.
Transport
Vibration test to EN 60068-2-6:1996,
shock test to EN 60068-2-27:1995
In its original packaging, the gas analyzer will withstand normal
shipping conditions.
Installation site requirements
Installation location
The gas analyzer is intended for indoor installation only. Installation location altitude max. 2000 m above sea level (over 2000 m
on request)
Ambient temperature
Operation: +5 to +45 °C, Uras26 in combination with another analyzer, Limas23, Fidas24: +5 to +40 °C, Storage and transport:
−25 to +65 °C
Relative humidity
< 75 %, slight condensation allowed
Air circulation
For sufficient air circulation, multiple housings in a 19-inch rack
must be installed with a separation of at least one height unit between housings.
Power supply
Input voltage
AC 100 to 240 V − 15/+ 10 %, 50 to 60 Hz ± 3 Hz
Power consumption
Max. 187 VA
Connection
3-pin plug per EN 60320-1/C14; connection cable supplied.
Connection diagrams see page 24.
Individual analyzers
Input voltage
DC 24 V ± 5 %
Power consumption
Uras26: max. 95 W Magnos27: max. 35 W
Limas23: max. 100 W ZO23: max. 35 W
Magnos206: max. 50 W Caldos27: max. 17 W
Magnos28: max. 50 W Fidas24: max. 40 W
Fidas24: Heating of detector and sample gas inlet
Input voltage
AC 115 V or 230 V, ± 15 % (max. 250 V), 47 to 63 Hz
Power consumption
125 VA for detector heating,
125 VA for heated sample gas inlet (option)
Connection
4-pin plug; connection cable supplied.
Connection diagrams see page 35.
Note regarding the analyzers performance characteristics
The performance characteristics of the analyzers have been determined according to IEC 61207-1:2010 “Expression of performance of gas analyzers – Part 1: General”. They are based on operation at atmospheric pressure (1013 hPa) and nitrogen as the
associated gas. Compliance with these characteristics when
measuring other gas mixtures can only be assured if their composition is known. The physical detection limit is the lower limit
of the performance characteristics relative to the measuring
range span.