and velocity sensors)
• CVxxx and VExxx vibration sensors (velocity sensors)
• CPxxx dynamic pressure sensors (piezoelectric pressure sensors)
• TQxxx proximity sensors
• LSxxx air-gap sensors.
Dynamic inputs
Number of channels : 4 (independent channels)
Voltage inputs
• DC measurement range : 0 to +20 VDC or 0 to −20 VDC.
Note: 10 Hz DC filter (see DC filtering on page 5).
• AC measurement range : ±20 VPEAK-PEAK
• AC + DC measurement range : ±24 VPEAK-PEAK
Common-mode voltage range : −50 to +50 VDC
Common-mode rejection ratio (CMRR) : >55 dB, up to 60 Hz.
>60 dB, from 45 to 65 Hz.
Current inputs
• DC measurement range : 0 to 35 mA
• AC measurement range : ±30 mAPEAK-PEAK
• AC + DC measurement range : ±50 mAPEAK-PEAK
Frequency bandwidth : DC to 20 kHz
Input impedance
• Voltage : ≥100 kΩ, between the differential (high and low) inputs
• Current : 200 Ω ±0.2%
Accuracy
• Amplitude : ±1% of full scale
• Phase : ±1° from 10 Hz to 2 kHz.
±15° from 2 to 20 kHz.
Dynamic input range : ≥80 dB, from 3 Hz to 20 kHz
DC filtering
DC filter
• Cutoff frequency (−3 dB) : 10 Hz ±3.5 Hz
• Roll-off : −40 dB/decade (second order)
Note: The DC filter is used to extract the DC part of a dynamic input when it is configured as a DC input.
High-pass filtering
High-pass filter
• Cutoff frequency (−3 dB) : 0.1, 1 or 3 Hz (or bypassed)
• Roll-off : −20 dB/decade (first order)
• Phase error : <1° at 100 times the cutoff frequency (10, 100 or 300 Hz)
Note: The high-pass filter is used to configure a dynamic input for an AC only input signal with one of 3 different
cutoff frequencies. This filter can be disabled in order to allow the DC-coupling of the input signal (AC + DC)
Auxiliary inputs
Number of channels : 2 (independent channels)
configurable as either tachometer inputs or DC inputs
Common-mode voltage range : −50 to +50 VDC
Common-mode rejection ratio (CMRR) : >50 dB, up to 60 Hz.
>55 dB, from 45 to 65 Hz.
Tachometer input
• Triggering method : Crossing of threshold on rising edge or falling edge of signal
• Triggering threshold : 2/3 of peak-peak value ±10% for rising edge.
1/3 of peak-peak value ±10% for falling edge.
• Tachometer range (on input) : 2 Hz to 50 kHz
• Speed / frequency measurement
range
: 1 to 65535 RPM / 1 Hz to 1092 Hz.
Note: After division by number of wheel teeth (1 to 255).
• Voltage range : 0.6 to 50 VPEAK-PEAK from 2 Hz to 10 kHz.
2 to 50 VPEAK-PEAK from 10 kHz to 50 kHz.
Auxiliary input
• Current range input : ±50 mAPEAK-PEAK (AC + DC measurement range)
• Voltage range input : ±50 VPEAK-PEAK
DC input
• Voltage measurement range : 0 to +20 VDC or 0 to −20 VDC.
Note: 10 Hz DC filter (see DC filtering on page 6).
• Current measurement range : ±50 mAPEAK-PEAK (AC + DC input)
Input impedance
• Voltage : ≥100 kΩ, between the differential (high and low) inputs
• Current : 200 Ω ±0.2%
Dynamic input range : ≥72 dB
DC filtering
DC filter
• Cutoff frequency (−3 dB) : 10 Hz ±3.5 Hz
• Roll-off : −40 dB/decade (second order)
Note: The DC filter is used to extract the DC part of an auxiliary input when it is configured as a DC input.
Sensor/measurement chain OK check
Number of levels : Up to 16 configurable threshold levels (16 DC regions)
OK level range
• Voltage inputs : ±20 VDC
• Current inputs : 0 to 23 mA
Operating principle
• SIL safety sensors : Line-fault detection of conditions such as a problem with the sensor
and/or cabling, problem with the signal conditioner, and/or other
problem with the measurement chain or power supply.
Note: Requires a SIL safety sensor/measurement chain that
provides a suitable diagnostic signal (DC bias level), for example,
measurement chains using IPC707 or IQS900 signal conditioners.
• Standard sensors : Powered sensors – line-fault detection of conditions such as
open-circuit or short-circuit.
Unpowered sensors – line-fault detection of conditions such as
open-circuit.
Digital signal processing
Analogue to digital converter (ADC) : 24 bit
Dynamic range : ≥80 dB
Frequency bandwidth : 0 Hz to 20 kHz
Accuracy
• Amplitude : ≤1% of input full scale
• Phase : ≤1.5°
Digital filtering
• Notch filter : 50 or 60 Hz
• ISO 2954 filter : 10 Hz to 1 kHz (−3 dB), −24 dB/octave
• Band-pass filter : <0.1 dB ripple in pass band, >55 dB attenuation in stop band, 0.1 or
3 dB attenuation at cutoff, −24 to −60 dB/octave slope
• High-pass filter : 0.25 to 400 Hz
• Low-pass filter : 10 Hz to 20 kHz
Measurement resolution : 2048 point waveform / 800 line spectrum
FFT window : Hanning
FFT resolution : 800 spectral lines
Integration count : 0, 1 or 2
Qualifiers (rectifiers) : RMS, Peak, Peak-Peak and Average.
Scaled Peak, Scaled Peak-Peak and Scaled Average.
Extracted data (measurements) : 2 to 10 processed outputs per channel/processing function.
See Processing functions on page 7.
Extracted data type : Scalar, Vector
Order tracking : Digital resampling
Update rate (internal) : 20 ms min. for time domain processing.
100 ms min. for frequency domain processing.
VibroSight® software update rate
(external)
: Configurable as 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 50 s, …
Processing functions
The following configurable signal processing blocks and measurements are supported by the