Advance Optima Module Uras 14
Physical principles
Measurement
principle
• NDIR Technique
(Non-Dispersive InfraRed Analysis)
• The measurement effect is based on resonance absorption of gas-specific
vibration-rotation bands of gas molecules with differing atoms in the median
infrared spectrum at wavelengths between 2.8 and 8 µm.
• The individual gases to be measured are identified by their specific absorption
bands. Each gas has such an absorption spectrum (fingerprint).
Exceptions:- Monoatomic gases, such as inert gases- Symmetrical gases, such as N2. O2 and H2- These types of gases cannot be measured with this method.
The relationship between measured infrared emission absorption and the sample
component is based on the LAMBERT-BEER law:
A = (I0 - I1) / I0 = 1 - e -ε(λ)⋅ρ⋅l
where
A
I0
I1
ε(λ)
ρ
l
= Absorption
= Emission entering the cell
= Emission leaving the cell
= Sample component extinction coefficient
= Sample component density
= Sample cell length
The relationship between test component density ρ and its volumetric
concentration c is
ρ = ρ0 ⋅ c ⋅ p/p0 ⋅ T0/T
where
ρ0
p0
T0
= Pure gas density
= Pressure
= Temperature
under standard conditions (1013 hPa, 0°C).
The second equation shows that the sample component's volumetric concentration
depends on the sample cell pressure and temperature.
The first equation finds a non-linear relationship between absorption and volumetric
concentration
Physical principles, continued
IR emission
Choppers
Sample cell
Infrared detector
• Generated by broad-band emitter
• Emitted as a beam package alternately in the form of a sample and reference
beam through the sample and reference chambers of the sample cell and is
partially absorbed by the sample component molecules
• Counterphase modulation by means of a motorized chopper wheel
• Both modulated beam packages appear alternately at the infrared detector
• Created by applicable regulation of the sample and reference beam balance
• Depending on the application, the sample chamber receives a sample, zero-point
or end-point gas flow so that a part of the infrared radiation is absorbed in a
concentration-dependent manner.
• The emission passes unhindered if the reference chamber is filled with a gas that
does not absorb infrared (N2).
• A two-part transmission detector with front and rear chambers filled with the gas
components to be measured; selectivity is determined by the infrared detector.
The two chambers are separated by an infrared-transparent window. Additionally,
the two chambers are separated by a stressed metal membrane with
counterelectrodes. This unit is known as the diaphragm capacitor.
•It reacts in the following manner in the presence of the sample component:
• IR radiation is weakened in the sample cell's sample chamber and enters the
receiver's front chamber.
• The equilibrium between the sample and reference beams initially established
by calibration and the aperture is now disturbed.
• There is an energy difference (temperature change) in the form of reduced
pressure in the front chamber.
• This pressure reduction is transformed into a capacitance change in the
membrane capacitor by deflecting the metal diaphragm.
• Since the diaphragm capacitor is connected to a high-impedance DC voltage, a
corresponding periodic AC signal is generated.
Determination of influence values
Associated gas
effects
Pressure
Flow rate
Temperature
The sample gas is a mixture of the sample component and associated gas
components. If the infrared absorption bands of one or more associated gas
components overlap the sample component's bands, the test results will be
affected.
The influence of interfering gas components is termed cross sensitivity or carrier
gas dependence.
Cross sensitivity is determined by connecting an inert gas (e.g. N2) which is mixed
with the interfering gas components (corresponding to the test gas).
The influence acts on the zero-point measurement value indication.
Carrier gas dependence, which is rarely observed, occurs when the physical
properties of the sample gas differ markedly from those of the test gas. This
interference changes the slope of the device's characteristic curve. This curve is
corrected at the end-point.
The Uras 14 has the following methods available for interference correction:
• Interference filter
• Filter cells
• Internal electronic cross-sensitivity correction
• Internal electronic carrier gas correction
According to the gas laws, the sample cell's volumetric concentration depends on
the pressure in the sample cell and is thus dependent on the process gas and air
pressure. This effect acts on the end-point and amounts to approx. 1% of the
measurement value per 1% of pressure change (therefore, per 10 hPa).
An internal pressure sensor reduces this effect to 0.2%.
The flow rate affects pressure in the sample cell and the module's T90 times.
The flow rate should be between 20 and 100 liters/hour.
Temperature has a markedly different effect on all optical components in the beam
path. This effect is reduced by:
• Temperature compensation
A temperature sensor in the first infrared detector's preamplifier measures the
temperature in the module.
This signal is used for electronic correction.
Zero-point effect: ≤ 1% of the measurement range per 10°C
End-point effect: ≤ 3% of the measured value per 10°C
• Thermostat (optional)
Zero-point effect: ≤ 1% of the measurement range per 10°C
End-point effect: ≤ 1% of the measured value per 10°C
Depending on the measurement application, the Uras 14 Analyzer can be
equipped with the following main components:
• 1 to 4 infrared detectors
• 1 to 2 beam paths
• Up to 2 infrared detectors per beam path
• The following elements are permanently installed
•Both emitter inserts are filled
•There is hardware support for installation of a thermostat
•IR module circuit board
•Sensor-CPU circuit board
•Pressure sensor circuit board
• Other components are fitted according to the measurement application or
configuration ordered.
• Any version of the module can be installed in a 19" rack or wall housing without
special conversion.
• The pneumatics module and oxygen analyzer module can be incorporated
together in the gas path.
Special components The following components can be fitted according to the option ordered or
measurement task to be carried out:
• 1 to 2 calibration units
• 1 to 2 filter cells
• Optics filter
• Gas paths
• FPM hose
• PTFE hose
• Stainless steel pipe